albumin reagent / for clinical chemistry / serum / colorimetric
A502-480

{{requestButtons}}

Characteristics

  • Type:

    albumin

  • Applications:

    for clinical chemistry

  • Sample type:

    serum

  • Analysis mode:

    colorimetric

Description

Test:
Albumin Reagent Set (A502-480)

Number of tests:
480 tests

Format:
Liquid

Method:
Endpoint BCG

Testing Procedure:
Manual & Automated

Storage Temperature:
Reagent is stored at room temperature (15-30°C)
Albumin standard is stored refrigerated (2-8°C)

Wavelength:
630 nm

Linearity:
0.5 – 8.0 g/dL

Expected Values:
3.5 – 5.3 g/dL

Reagent Deterioration:
The reagent should be a clear, yellow-green solution. If turbidity or precipitation has
occurred, the reagent should be discarded.

Limitations of Procedure:
Ampicillin and other medications seriously interfere with the dye-binding properties of
albumin.
The dye-binding properties of albumin, other than human, differ among species.
Sample with values above 8.0 g/dL should be diluted with 0.9% saline 1:1, re-run, and
results multiplied by 2. Samples with results below 0.5 g/dL should be done
electrophoretically.

Other Teco Diagnostics products

Clinical Chemistry