OverviewThe SEEG 100 is an EEG performance tester engineered for regulatory verification, R&D and production-line validation. It supports IEC 60601-2-26:2012, GB 9706.226-2021 and related standards and delivers sub-microvolt signal generation with precise timing and amplitude control.
Highlights- Compliance coverage: IEC 60601-2-26:2012, GB 9706.226-2021, JJG 1043, JJF 1388, JJF 1390
- High sensitivity output capable of 1 µV signals with sophisticated shielding and electrical design
- 9 channels with auto-select switches for individual channel testing (output leads Ch1–Ch9)
- Built-in test circuits for input impedance checks with selectable 620kΩ/4.7nF
- Playback and load support for EDF (PhysioNet), binary and text formats
- Integrated test sequences, options and pass criteria to simplify standard-compliance testing
- SDK available for rapid development of custom or automated test software
NoteTo enhance product appearance consistency, WhaleTeq updated the exterior printing of the SEEG 100 on July 28, 2023. Functional features remain unchanged.
Package contents (standard)- SEEG 100 x 1
- EEG breakout box x 1
- Wire tie x 22
- USB cable x 1
- Grounding wire x 1
Technical Specifications (summary)Main output voltage accuracy: ±1% for amplitudes ≥50µVpp
Main output voltage resolution (DAC): 0.5µV
Frequency accuracy: ±0.1%
Pulse timing accuracy: ±0.2 ms
Resistor tolerance: ±0.5%
Capacitor tolerance: ±5%
Precision 1000:1 divider (100kΩ:100Ω): ±0.05%
Sample rate: 5 kHz ±0.05%
DC offset (fixed, from internal supercapacitor): 300 mV ±0.1%
DC offset (variable): setting ±1% or ±3 mV; may include up to 50 µVpp noise
Power: typical load <0.25 A; up to 0.45 A if all relays engaged
Operating conditions: 5–40°C, 50–80% RH, altitude up to 2000 m
Safety and signal integrity: USB IC protection and dedicated filters for microcontroller and DC/DC noise reduction
Signal Types (settings summary)Sine: 0.05–500 Hz; default 10 Hz; step 0.01 Hz
Sine amplitude: -2000–2000 µV; default 100 µV; step 1 µV
Triangle: 0.05–500 Hz; default 10 Hz; step 0.01 Hz — amplitude -2000–2000 µV; step 1 µV
Square: 0.05–500 Hz; default 10 Hz; step 0.01 Hz — amplitude -2000–2000 µV; step 1 µV
Rectangle pulse: 0.05–5 Hz; pulse width 2–300 ms; default width 100 ms
Triangle pulse: 0.05–5 Hz; pulse width 2–300 ms; default width 100 ms
Hysteresis signals: 0.05–500 Hz; amplitude -2000–2000 µV; pulse width 2–300 ms
Signal Add-onDC offset: -1000–1000 mV; step 1 mV
Input impedance selection 620kΩ / 4.7nF: on/off (default off)
Main noise presets: 50 Hz (default), 60 Hz, 80 Hz, 100 Hz (80/100 Hz for capacitance correction only)
Frequency scan (sine): start 0.67–500 Hz; stop 0.67–500 Hz; duration 10–180 s
Output lead electrode: Ch1–Ch9 (Ch10–Ch22 shorted); default Ch1
Documentation & Software (available)- Application / Software package (AP)
- Datasheet
- User manual
- SDK and release notes
Technical characteristics- Model: SEEG 100
- Standards: IEC 60601-2-26:2012, GB 9706.226-2021, JJG 1043, JJF 1388, JJF 1390
- Channels: 9 with auto-select switches
- Signal amplitude range: -2000 to 2000 µV
- Sample rate: 5 kHz
- DAC resolution: 0.5 µV
- Precision divider: 1000:1 ±0.05%
- Input impedance test option: 620kΩ / 4.7nF selectable
- File support: EDF (PhysioNet), binary and text formats
- SDK available for automation and custom test solutions