Overview
The MicroScan autoSCAN-4 System is a semi-automated clinical microbiology instrument for bacterial identification (ID) and antimicrobial susceptibility testing (AST). Designed for use as an adjunct in high-complexity labs or as a primary analyzer in low-volume settings, it processes conventional and specialty ID/AST panels with rapid, standardized readings to support diagnostic workflows.

Key features
  • Semi-automated operation with minimal training required for laboratory staff
  • Processes conventional panels using direct, identification-independent MICs to aid detection of emerging resistance patterns
  • Runs specialty ID panels for short turnaround when rapid identification is required
  • Optional remote diagnostic service to support uptime and troubleshooting
  • Comprehensive service and technical support for clinical microbiology applications


Software
Supplied with LabPro Information Manager software to streamline laboratory workflow and reduce operator interaction. LabPro Alert EX automatically flags atypical or critical results to assist rapid reporting and clinical decision-making.

Specifications (table)
Physical:
Height: 25.4 cm (10 in); Width: 48.3 cm (19 in); Depth: 58.4 cm (23 in)
Weight: 18.1 kg (40 lb)
Power cord length: 16.5 cm (6.5 in)
Computer connection cable length: 21.6 cm (8.5 in)
Digital interface: RS-232C (9,600 Baud)
AC plug: 5-15P

Power:
Line voltage: 105/115/230 V AC ±10%
Line frequency: 60/50 Hz ±10%
Line current: 2.00 A max. (115 V), 1.25 A max. (230 V)
Heat dissipation: 630 BTU/h
Uninterruptible Power Supply (UPS): Optional

Optical system:
Shortened filter photometer with six interference filters (wavelengths nm: 440, 470, 505, 560, 590, 620), 96-channel fiber optic cable
Light source: Tungsten-halogen lamp, ~3,500 hours life
Spectral bandwidth: 10 nm

Analysis system:
Panel capacity on system: 1 panel/reading
Read time: Less than five seconds

Safety requirements:
Earth impedance: 0.1 Ohm max.
Dielectric strength: No dielectric breakdown (at 1,000 V AC and 60 Hz for one minute)
Power cord: 3-conductor, 16 AWG, industrial quality, SJT
Overcurrent protector: 1.60 A (115 V), 0.80 A (220 V)
Leakage current: Less than 500 µA

Regulatory compliance and environment:
Designed and tested to UL 3101-1, CSA C22.2 No. 1010.1 and EN 61010-1 standards under specified environmental conditions
Operating temperature: 5 °C (41 °F) to 40 °C (104 °F)
Humidity: Maximum 80% at 31 °C (88 °F) to 50% at 40 °C (104 °F)
Operating altitude: Up to 2,000 m (6,561 ft)
Main power: 115/230 V AC, 50/60 Hz, 230 W max.
Overvoltage category: Category II, connection to branch circuit
Pollution degree: Degree 2 (normal indoor laboratory environment)
Emissions: Developed and tested to EN 61326

Characteristics / Technical specifications
  • Product name: MicroScan autoSCAN-4 System
  • Model family: autoSCAN-4
  • Typical panel processing: conventional ID/AST panels and specialty ID panels
  • Panel read capacity: 1 panel per measurement; read time < 5 seconds
  • Optical detection: 6 interference filter photometer (440, 470, 505, 560, 590, 620 nm)
  • Light source life: approx. 3,500 hours (tungsten-halogen)
  • Interfaces: RS-232C serial connection
  • Power: 105/115/230 V AC ±10%, 50/60 Hz; max power 230 W
  • Weight & dimensions: 18.1 kg; H 25.4 cm x W 48.3 cm x D 58.4 cm
  • Operating conditions: 5–40 °C, up to 80% RH (at lower temps), altitude to 2,000 m
  • Safety and standards: UL 3101-1, CSA C22.2 No. 1010.1, EN 61010-1; emissions per EN 61326

Catalogs