Non-destructive testing, specially designed for lab use as well as inline testing
Deterministic
Quantifiable and repeatable results according to USP standard
Versatile
Turnkey solution with customized test chambers, adaptable to vials, syringes, blister, intravenous (IV) bags and other non-porous packages
Technical specifications
Sensitivity - Down to 0.4 µm, 2 · 10⁻⁵ mbar · l/s
CDA supply - required for operation
CDA quality - 1.3.1 according to ISO 8573-1
CDA pressure (bar rel.) - 6.3 - 10 bar rel.
CDA pressure (psig) - 91 - 145 psig
Typical CDA consumption - 2 Nl/cycle
Calibration gas supply - optional
Calibration gas pressure (bar rel.) - 6.3 - 10 bar rel.
Calibration gas pressure (psig) - 91 - 145 psig
Venting gas - optional
Venting gas pressure (bar rel.) - 0 - 1.5 bar rel.
Venting gas pressure (psig) - 0 - 22 psig
Power supply - 90 - 250 V AC
Typical power consumption - 1200 W
Operating temperature - 20 - 25 °C
Humidity - 30 - 65 %
User interface - 10" multi-touch full HD color screen
Software - 21 CFR part 11 compliance, PDF GMP test and calibration reports, authentication local or domain (LDAP), optional remote access to data
Operating system - Windows 10
Network connection - 1x LAN (RJ45) WiFi AP
Interfaces - 2x USB 3.0 (ext.), 1x HDMI
Sound pressure level - <53 dB(A)
Weight approx. - 130 kg
Dimensions (L x W x H) (mm) - 913 x 558 x 1071 mm
Dimensions (L x W x H) (inch) - 913 x 558 x 1071 inch
Accessories
Customized test chambers
Calibration gas supply
Venting gas
Possibility to use a customer laptop via Ethernet or WiFi connection
Remote access to data