OverviewClass I microbiological safety cabinets provide operator protection by controlling inward airflow through the front aperture and filtering exhaust air. The Bio 1+ range from Envair is designed for laboratory containment tasks that require Class I protection and is suitable for handling biological agents up to ACDP Hazard Group 3 (site-dependent for certain agents such as Covid-19).
Product summaryThe Envair Bio 1+ cabinets are compact, stainless-steel Class I units available in multiple sizes and configurations. They are offered with either recirculating or ducted exhaust and can incorporate integral fumigation systems for decontamination between uses.
Options- Recirculating or ducted exhaust configurations
- Optional recirculating Formonia fumigation system
- Compliance with BS EN 12469:2000
- Variety of sizes, configurations and fumigation/accessory options
- Designed for easy cleaning and maintenance
- Quick installation suitable for temporary sites
Key features- Stainless-steel construction with white polyester coating and polished stainless-steel work base
- Smooth internal/external surfaces and radiused internal corners for effective cleaning
- High containment performance delivering operator protection
- Integral double HEPA filtration on recirculating models, filters to H14 (BS EN 1822)
- Replaceable pre-filter to extend HEPA life
- Exhaust fan box with integral fans
- Hinged toughened glass front visor supported by gas struts
- Self-sealing lightweight night door/closure panel with interlock for safe fumigation
- Comprehensive alarms and interlocks for operational safety
- Membrane control panel with clear indicators and alarm functions
- Automatic speed-controlled fan for optimal filter life and consistent performance
- Front-access design for rapid setup and maintenance
- BMS and auto-damper signal outputs for facility integration
Technical specifications- Protection principle: inward airflow through front aperture with exhaust filtration (Class I)
- Intended use: handling biological agents up to ACDP Hazard Group 3 (site-dependent for certain agents such as Covid-19)
- Standards: compliant with BS EN 12469:2000; HEPA filters to H14 (BS EN 1822)
- Filtration: integral double HEPA on recirculating models; replaceable pre-filter
- Exhaust: recirculating or ducted exhaust options; exhaust fan box with integral fans
- Fumigation: optional Formonia recirculating fumigation system; self-sealing night door with interlock for safe fumigation
- Construction: stainless steel with white polyester coating; polished stainless-steel work base; smooth surfaces and radiused internal corners
- Access & visibility: hinged toughened glass front visor supported by gas struts
- Controls & safety: membrane control panel, comprehensive alarms and interlocks, automatic speed-controlled fan
- Integration: BMS and auto-damper signal outputs
- Serviceability: front access for rapid set-up and maintenance