OverviewCompact tabletop steam sterilizer designed for medical, dental and laboratory use. The DGT-B Series delivers Class B sterilization cycles with a built-in steam generator, automatic door and microcomputer control for repeatable, validated sterilization processes.
Key features- Automatic door switch for safe loading/unloading
- Automatic water feed and built-in steam generator for rapid saturation
- Stainless steel inner chamber and sealed cover
- Microcomputer control with LED display and capacitive/inductive operation
- Drying function and automatic cold air discharge/steam exhaust after cycle
- Safety interlocks and multiple protections: over-temperature, over-pressure, low water level, anti-dry burn, door lock
- Three-stage pre-vacuum for effective sterilization of wrapped instruments
- Supports BD test, HELIX test and Vacuum test for process validation
- Built-in printer, USB interface and U-disk support for cycle records
- European CLASS B standard compliance
Technical parametersModel | DGT-18B | DGT-23B | DGT-45B
Volume | 18 L | 23 L | 45 L
Power | 2 kW | 2 kW | 2.5 kW
Voltage | 110V / 220V, 60Hz / 50Hz | 110V / 220V, 60Hz / 50Hz | 110V / 220V, 60Hz / 50Hz
Design pressure | 0.25 MPa | 0.25 MPa | 0.25 MPa
Design temperature | 139 °C | 139 °C | 139 °C
Rated working pressure | 0.22 MPa | 0.22 MPa | 0.22 MPa
Rated working temperature | 134 °C | 134 °C | 134 °C
Chamber dimension (Φ×L, mm) | Φ250×370 | Φ250×470 | Φ316×615
Outer dimension (L×W×H, mm) | 620×505×450 | 505×710×450 | 657×870×575
Packing size (L×W×H, mm) | 670×560×510 | 760×560×510 | 980×750×630
Gross / Net weight | 62 / 53 kg | 65 / 58 kg | 142 / 105 kg
Safety & standards- Compliant with European CLASS B sterilization requirements
- Multiple automatic protections to prevent overheating, overpressure and dry burning
- Safety interlocking device and door lock to prevent accidental opening during cycle
- Audible buzzer and automatic stop upon cycle completion
Operation, validation & data- Microcomputer-controlled cycles with LED display and intuitive operation
- Supports BD, HELIX and vacuum tests for routine validation
- Integrated printer plus USB/U-disk for export and storage of cycle records
- Equipped with standard test interface for external validation tools
Applications- Designed for clinics, dental practices, medical and research laboratories, veterinary clinics and other professional environments requiring validated Class B sterilization.