Intended useTreatment of acute or chronic renal failure by extracorporeal hemodialysis to remove metabolic waste (urea, creatinine), β2-microglobulin, phosphate and excess fluid.
Key features- Polyethersulfone (PES) hollow-fiber membrane designed for high-flux clearance of small, middle and large molecules
- Good biocompatibility
- Continuous toxin clearance capacity with low albumin loss
- Housing and potting materials engineered for secure assembly and safe use
- Available in multiple surface areas and chamber volumes to match clinical prescriptions
Specifications- Models: OCI-HD13M, OCI-HD15M, OCI-HD16M, OCI-HD17M, OCI-HD18M, OCI-HD19M, OCI-HD20M, OCI-HD21M, OCI-HD22M, OCI-HD23M, OCI-HD24M, OCI-HD25M
- Surface area: 1.3–2.5 m² depending on model
- Ultrafiltration coefficient (mL/h·mmHg): 39–77 depending on model
- Chamber volume: 84–146 mL depending on model
- Membrane material: Polyethersulfone (PES)
- Housing material: Polycarbonate (PP)
- Potting compound: Polyurethane (PU)
- Test conditions: T = 37℃, QD = 500 mL/min, QF = 10 mL/min (in vitro simulation)
Performance (solute clearance)- Urea (mL/min): QB 200: ~180–199; QB 300: ~259–295; QB 400: ~292–364
- Creatinine (mL/min): QB 200: ~164–190; QB 300: ~239–270; QB 400: ~259–295
- Phosphate (mL/min): QB 200: ~164–203; QB 300: ~244–277; QB 400: ~284–324
- Vitamin B12 (mL/min): QB 200: ~129–162; QB 300: ~159–193; QB 400: ~168–216
Screening coefficients & operational ranges- Albumin screening coefficient: ≤ 0.01
- β2-microglobulin screening coefficient: 0.8 ± 0.2
- Inulin screening coefficient: 0.9 ± 0.1
- Blood flow range: 200–400 mL/min
- Dialysate flow range: 500–800 mL/min
- Maximum TMP: 66.5 kPa / 500 mmHg
Sterilization & shelf life- Irradiation sterilization; shelf life: 3 years