laboratory climate chamber / temperature / humidity / 1-door
HPP110

Add to MyMedicalExpo favorites
Compare this product Don’t compare this product
laboratory climate chamber / temperature / humidity / 1-door laboratory climate chamber / temperature / humidity / 1-door - HPP110
{{requestButtons}}

Characteristics

  • Applications:

    laboratory

  • Tested parameter:

    temperature, humidity

  • Number of doors:

    1-door

  • Other characteristics:

    stainless steel, constant

  • Temperature range:

    Min.: 0 °C (32 °F)

    Max.: 70 °C (158 °F)

Description

Constant climate chamber HPP110 Our constant climate chamber is tailored to environmental simulation, material testing and stability testing in accordance with the ICH guidelines.

Constant climate chamber HPP110
Standard delivery Constant climate chamber
On this page, you can find all the essential technical data on the Memmert stability chamber HPP. Our customer relations team will be pleased to help if you want further information. If you should require a customised special solution, please contact our technical specialists at myAtmoSAFE@memmert.com.

Sketch Constant climate chamber HPP110
Temperature
Setting temperature rangewithout light, with humidity: +5 to +70 °C
Setting temperature rangewith light, with humdity: +5 to +70 °C
Setting temperature rangewith light, without humditiy: 0 to +70 °C
Setting temperature rangewithout light, without humidity: 0 to +70 °C
Working temperature rangewith light, without or with humidity: 15 to +40 °C
Working temperature rangewithout light, with humidity: +5 (at least 20 below ambient temperature) to +70 °C
Working temperature rangewithout light, without humidity: 0 (at least 20 below ambient temperature) to +70 °C
Setting accuracy temperature0.1 °C
Temperature sensor2 Pt100 sensors DIN Class A in 4-wire-circuit for mutual monitoring, taking over functions in case of an error
Humidity
Setting range humidity10 - 90 % rh
Humidityhumidity supply with distilled water from external tank by self-priming pump
Humidificationhumidification by hot steam generator
Dehumidificationdehumidification by cold trap using the Peltier technology
Setting accuracy humidity0.5 % rh