Product overviewCompact tube furnaces of the RT series for laboratory and process applications requiring horizontal, vertical or angled operation. Adjustable tilt angle (0–90°) and working height allow flexible positioning; compact design enables integration into existing process systems. Models include RT 50/250/11, RT 50/250/13 and RT 30/200/15 with Tmax options up to 1500 °C.
Standard equipment- Tmax 1100 °C, 1300 °C or 1500 °C
- Compact dual‑shell ventilated housing in textured stainless steel for low external surface temperature
- Insulation using non‑classified materials per EC Reg. No 1272/2008 (no RCF)
- Vertical or horizontal operation infinitely adjustable; angle 0–90°
- Working height infinitely adjustable
- Operation possible without stand if safety guidelines are observed
- Ceramic working tube C 530 including two fiber plugs for operation under air
- Type S thermocouple
- Low‑noise operation using solid state relays
- Heating wires wound directly around the working tube for 1100 °C and 1300 °C models (very fast heat‑up)
- Radial heating elements around the working tube for 1500 °C models (facilitates tube exchange)
- Control system integrated in furnace base; controller B510 with touch operation (5 programs × 4 segments)
- NTLog Basic for process data recording via USB flash drive
- Defined application within the constraints of the operating instructions
Additional equipment- Over‑temperature limiter with adjustable cutout temperature to protect furnace and load
- Gas supply system 1 for non‑flammable process gases (e.g. N2, Ar, He, CO2, forming gas)
- Process control and documentation via VCD software package for monitoring, documentation and control
Technical specifications (table)Model | Tmax (°C) | Outer dimensions W x D x H (mm) | Inner tube Ø (mm) | Heated length (mm) | Length constant temperature ±5 K (mm) | Tube length (mm) | Connected load (kW) | Connection | Weight (kg)
RT 50/250/11 | 1100 | 350 x 380 x 740 | Ø 50 | 250 | 80 | 360 | 2 | 1‑phase | 25
RT 50/250/13 | 1300 | 350 x 380 x 740 | Ø 50 | 250 | 80 | 360 | 2 | 1‑phase | 25
RT 30/200/15 | 1500 | 445 x 475 x 740 | Ø 30 | 200 | 70 | 360 | 2 | 1‑phase | 45
Notes1) Values refer to outside of the tube. Difference to temperature inside the tube up to +50 K. 2) External dimensions vary with additional equipment; dimensions on request. 3) Connected load refers to the standard furnace and may increase with additional equipment; for multi‑range voltage options the connected load applies to the highest permissible voltage.
Characteristics / technical specifications- Tmax options: 1100 °C, 1300 °C, 1500 °C
- Infinitely adjustable vertical or horizontal operation, angle 0–90°; working height adjustable
- Ceramic working tube C 530 with two fiber plugs for air operation
- Thermocouple type S; controller B510 (touch) and NTLog Basic data logging
- Heating concept: wires wound around tube (1100/1300 °C) or radial elements (1500 °C) for easy tube exchange
- Dual‑shell ventilated stainless steel housing; insulation without classified materials (no RCF)
- Examples: RT 50/250/13 — inner Ø 50 mm, heated length 250 mm, ~2 kW, ~25 kg