The SILBERMANN LEO-MN Manifold incorporates the basic necessities for distributing and monitoring medical gas throughout a facility.
This automatic changeover unit is designed for a continuous supply of gas at a pressure of up to 8 bar (higher outlet pressure upon customer request) and includes a mechanism that switches the supply from the "in service" bank to a "reserve" supply bank when the "in service" bank is empty.
FEATURES & BENEFITS- The SILBERMANN Automatic changeover unit is designed for a continuous supply of gas at a pressure of up to 8 bar (higher outlet pressure upon customer request).
- The panel includes a mechanism that switches the supply from the “in-service” bank to a “reserve” supply bank when the in-service” bank is empty.
- After rotating the lever and replacing the empty bank cylinders, the previous "in-service " becomes the “reserve” bank for the next operation.
- The panel is equipped with two manifold pressure regulators (in case of malfunction), a mid-range gauge, line pressure regulators w/bypass valves, all mounted on a white baked-painted steel panel.
- Safety valve @ 6 bar supplied with the panel.
TECHNICAL DATA- Panel finish – baked-on powder-coated.
- Piping System: All materials are oxygen-compatible.
- Changeover mechanism: pressure difference
- Output connections for local alarms:
- One set of dry contacts N.O
- Flow Capacity: Oxy/medical air (4 bar) - 90 m³/hr.
- High-pressure regulator inlet – G 3/4” Female nut.
- Max inlet (cylinder) pressure: 200 bar.
- Standard oxy. Outlet pressure: 4- 5 bar.
- Standard Outlet pressure for Nitrogen is 8 bar.
- Higher pressure is available upon request.