Product introductionGIT-SE01 is designed for in-situ testing of glove integrity on isolators and RABS. The device uses a pressure-decay method in accordance with ISO14644-7 to detect leaks in glove, sleeve and port assemblies, supporting routine and periodic integrity verification in aseptic environments.
Key capabilities- In-situ testing without glove removal.
- Automated test sequences compatible with cGMP workflows.
- Stainless steel construction suitable for cleanroom use.
- Integrated vacuum/pump unit.
- Built-in thermal printer for real-time records.
- Electronic records functionality compatible with 21 CFR Part 11.
- Test method aligned with ISO14644-7.
Scope / What it tests- Integrity testing of complete port, sleeve and glove assemblies.
- Leak detection for gloves and sleeves used in aseptic isolators and RABS.
- Intended for pre- and post-use testing as required by applicable regulations.
ComplianceThe GIT-SE01 implements pressure-decay testing consistent with ISO14644-7 Annex E.5 and supports recommendations from regulatory guidance including EU GMP Annex 1 (2022) clause 4.21 and FDA Guidance for Industry: Sterile Drug Products Produced by Aseptic Processing (2004) Appendix 1. The device supports routine physical integrity tests and periodic verification as recommended by these standards.
Applications- Pharmaceutical and biotechnology aseptic production
- Food industry safety control
- High-risk industrial protection
Technical specifications- Power supply: 220 V / 50 Hz
- Power: 200 W
- Input pressure: 2–6 bar
- Inflation/test pressure: 500–2,500 Pa
- Typical test time: 5–8 min
- Weight: 12.5 kg
- Detectable pore size: 300 μm
- Pressure resolution: 0.1 Pa
- Dimensions: 340 × 245 × 330 mm
- Detection port diameter: Ø255 mm (customizable)