Membrane ultrafilter Vivacon® 2 ETO
for liquidsfor protein solutionslaboratory

Membrane ultrafilter - Vivacon® 2 ETO - Vivaproducts, Inc. - for liquids / for protein solutions / laboratory
Membrane ultrafilter - Vivacon® 2 ETO - Vivaproducts, Inc. - for liquids / for protein solutions / laboratory
Add to favorites
Compare this product

Characteristics

Filter type
membrane
Fluid
for liquids, for protein solutions
Applications
laboratory, centrifuge, for protein concentration, for protein purification, for the pharmaceutical industry
Material
cellulose triacetate, polycarbonate, polypropylene, silicone, regenerated cellulose
Other characteristics
sterile
Connecting diameter

17 mm
(0.67 in)

Description

Product overview
Vivacon® 2 ETO centrifugal concentrators enable rapid DNA concentration and PCR primer removal for samples up to 2 ml. They use low-binding Hydrosart® regenerated cellulose membranes positioned horizontally for optimal recovery. Compatible with fixed-angle rotors that accept 15 ml centrifuge tubes. Reverse spinning into the upper recovery cap simplifies sample retrieval. Devices are sterilized by a dual-cycle ethylene oxide (ETO) gas process to a SAL of 10⁻⁶ and tested by qPCR (55 cycles) with a detection limit of 2 pg, ensuring no detectable human genomic DNA contamination.

Key features
  • Maximum starting volume: 2 ml.
  • Concentration factor: up to 30×.
  • Low-binding Hydrosart® membranes (various MWCO up to 100 kDa) and cellulose triacetate option (125 kDa).
  • Designed for use in fixed-angle rotors fitting 15 ml conical tubes; backspin recovery cap for simplified retrieval.
  • ETO sterilized (dual-cycle), PCR-grade.


Applications
  • DNA rebuffering of organic extraction samples prior to sequencing.
  • Dye removal after sequencing reactions.
  • Primer removal and plasmid concentration.
  • Concentration of nucleic acids and proteins for downstream analysis.


Performance - Vivacon® 2 for DNA
  • 2,000 MWCO — Time to concentrate up to 30×: 120 min at 20°C | Concentrate recovery: 92% | g-force: 7,500 xg
  • 10,000 MWCO — Time: 60 min at 20°C | Recovery: 94% | g-force: 5,000 xg
  • 30,000 MWCO — Time: 60 min at 20°C | Recovery: 95% | g-force: 2,500 xg
  • 50,000 MWCO — Time: 45 min at 20°C | Recovery: 96% | g-force: 2,500 xg
  • 100,000 MWCO — Time: 30 min at 20°C | Recovery: 93% | g-force: 2,500 xg
  • 125,000 MWCO (Cellulose Triacetate) — Time: 30 min at 20°C | Recovery: 88–89% | g-force: 2,000 xg


Performance - Vivacon® 2 for proteins
  • 2,000 MWCO — 0.25 mg/ml cytochrome c | Time to concentrate up to 30×: 120 min at 20°C | Recovery: 95% | g-force: 7,500 xg
  • 10,000 MWCO — 0.25 mg/ml cytochrome c | Time: 90 min | Recovery: 96% | g-force: 5,000 xg
  • 30,000 MWCO — 1.0 mg/ml BSA | Time: 40 min | Recovery: 96% | g-force: 5,000 xg
  • 50,000 MWCO — 1.0 mg/ml BSA | Time: 30 min | Recovery: 94% | g-force: 5,000 xg
  • 100,000 MWCO — 1.0 mg/ml bovine IgG | Time: 30 min | Recovery: 92% | g-force: 5,000 xg
  • 125,000 MWCO — 1.0 mg/ml bovine IgG | Time: 27 min | Recovery: 81% | g-force: 5,000 xg


Conversion table
  • Hydrosart® — 2 kDa MWCO — Double-stranded nucleotide cut-off: > 10 bp
  • Hydrosart® — 10 kDa MWCO — Cut-off: > 30 bp
  • Hydrosart® — 30 kDa MWCO — Cut-off: > 50 bp
  • Hydrosart® — 50 kDa MWCO — Cut-off: > 300 bp
  • Hydrosart® — 100 kDa MWCO — Cut-off: > 600 bp
  • Cellulose Triacetate — 125 kDa MWCO — Cut-off: > 650 bp


Ordering information
PCR Grade, 100,000 MWCO, Hydrosart® membrane, 100 per pack (Prod. #: VN02H42ETO) Price: $848.00

Specifications
  • Concentrator capacity (fixed-angle rotor): 2 ml.
  • Total length: 125 mm (concentration) / 115 mm (backspin); Width: 16 mm.
  • Active membrane area: 0.95 cm².
  • Hold-up volume (membrane and support): <10 µl.
  • Dead stop volume (fixed-angle rotor 25°): 55 µl; swing-bucket rotor: no dead stop.
  • Materials: Body – polycarbonate; Filtrate vessel – polypropylene; Backspin vial – polypropylene; Concentrator cap – polypropylene; O-ring – silicone.
  • Membrane types: Hydrosart® (various MWCO up to 100 kDa); Cellulose Triacetate (125 kDa).
  • Equipment required: Centrifuge with fixed-angle rotor fitting 15 ml (17 mm) conical bottom tubes (95 mm depth).
  • Maximum recommended speed: 2,500 xg (may be increased in some cases; DNA recovery can vary with g-force).

Catalogs

No catalogs are available for this product.

See all of Vivaproducts, Inc.‘s catalogs
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.