Intended useGlove box designed for BSL-4 containment and handling of hazardous biological agents. The enclosure is airtight and maintained under negative pressure. Autoclaves, centrifuges and other laboratory instruments can be integrated.
Key product points- Supply and exhaust fitted with HEPA filters.
- Internal negative pressure maintained at -130 Pa and continuously monitored.
- Available as standalone units or modular, connectable configurations.
Standards and specificationsModel: BHC-1301Ⅲ
Filtration efficiency: ≥99.99% for 0.3 μm particles (PAO)
Filter elements: Supply/Exhaust: HEPA filters (scan-tested items)
Internal pressure: Indicated by pressure gauge
Air exchange rate: ≥50 cycles/h
Leak tightness: ≤1×10−7 cc/sec at 75 mmAq (halogen scanning test on welds and joints under 75 mmAq pressure)
Construction: Body: SUS304 (interior corners R≥10 mm, continuous weld finish)
Frame: painted steel plate; Butterfly valves: sealed type (mounted at inlet/outlet)
Window: tempered or laminated glass; Gaskets: neoprene rubber
Lighting / sterilization lamp: 15W × 1 / 15W × 1
Dimensional drawingDimensional drawing available (product main unit drawings attached)
Features / Technical specifications (summary)- Model: BHC-1301Ⅲ
- Filtration efficiency: ≥99.99% for 0.3 μm particles
- Supply/Exhaust HEPA filters installed (scan-tested)
- Negative pressure maintenance: -130 Pa (monitored)
- Air exchange rate: ≥50 cycles/h
- Leak tightness: ≤1×10−7 cc/sec at 75 mmAq (halogen test)
- Body material: SUS304 (interior corner R≥10 mm, continuous weld)
- Frame: painted steel plate
- Window: tempered or laminated glass
- Gaskets: neoprene rubber
- Lighting / sterilization lamp: 15W × 1 / 15W × 1
- Compatibility: standalone or modular; autoclave and centrifuge integration possible