Decontamination transfer hatch PBL series
for cleanroomslaminar flowwith HEPA filter

Decontamination transfer hatch - PBL series - Tailin - for cleanrooms / laminar flow / with HEPA filter
Decontamination transfer hatch - PBL series - Tailin - for cleanrooms / laminar flow / with HEPA filter
Decontamination transfer hatch - PBL series - Tailin - for cleanrooms / laminar flow / with HEPA filter - image - 2
Decontamination transfer hatch - PBL series - Tailin - for cleanrooms / laminar flow / with HEPA filter - image - 3
Decontamination transfer hatch - PBL series - Tailin - for cleanrooms / laminar flow / with HEPA filter - image - 4
Decontamination transfer hatch - PBL series - Tailin - for cleanrooms / laminar flow / with HEPA filter - image - 5
Decontamination transfer hatch - PBL series - Tailin - for cleanrooms / laminar flow / with HEPA filter - image - 6
Decontamination transfer hatch - PBL series - Tailin - for cleanrooms / laminar flow / with HEPA filter - image - 7
Add to favorites
Compare this product

Characteristics

Options
for cleanrooms, decontamination, custom, hydrogen peroxide, with HEPA filter, laminar flow

Description

Introduction
The PBL Series Sterile Pass Box (decontamination airlock) is equipped with an integrated Vaporized Hydrogen Peroxide (VH2O2) system for surface decontamination. It features vertical laminar airflow and maintains a Class A internal environment. Controlled by a SIEMENS PLC, the system operates automatically and unattended once sterilization parameters are set. Different sizes can be customized.

Features
  • Using VH2O2 102/202 as a biological decontaminant, maintaining low temperature and normal pressure during the decontamination process for efficient and environmentally friendly operation.
  • Real-time monitoring of temperature, humidity, and pressure.
  • Provides a complete 4Q programme with on-site validation.
  • Meets FDA 21 CFR Part 11 requirements.
  • H14 HEPA filter system.
  • Pneumatic interlock door.
  • Pneumatic sealed door with toughened glass in stainless steel frame.
  • Vertical laminar flow, Class A interior.

Applications
  • Transferring materials such as APIs and tools from low grade areas to Class B areas.
  • Handling items that cannot endure high temperatures into Class A/B aseptic areas.

Technical specifications
  • Air flow: Vertical laminar flow
  • Cleanliness: GMP Grade A
  • Complete Cycle Time: 150 min
  • VH2O2 Residue: ≤1 ppm
  • HEPA Filters: H14
  • Outlet: 4" or 6" TC
  • VH2O2 Inlet: 1.5" TC
  • Sterility Assurance Level: ≥6-log kill
  • Chamber Material: SUS316L
  • Print: Bio-decontamination cycle parameter saving and printing
  • Alarms: High/low pressure alarm, interlock alarm, VH2O2 concentration alarm, temperature alarm, door open timeout

Catalogs

No catalogs are available for this product.

See all of Tailin‘s catalogs

Exhibitions

Meet this supplier at the following exhibition(s):

CPHI China
CPHI China

16-18 Jun 2026 Shanghai (China) Hall N3 - Stand B56

  • More information
    FILTECH
    FILTECH

    30 Jun - 02 Jul 2026 Cologne (Germany) Hall 7 - Stand N21

  • More information
    *Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.