Laboratory thermostat Universa U 420 P
benchtopdigitalelectronic

Laboratory thermostat - Universa U 420 P - LAUDA - benchtop / digital / electronic
Laboratory thermostat - Universa U 420 P - LAUDA - benchtop / digital / electronic
Laboratory thermostat - Universa U 420 P - LAUDA - benchtop / digital / electronic - image - 2
Laboratory thermostat - Universa U 420 P - LAUDA - benchtop / digital / electronic - image - 3
Add to favorites
Compare this product

Characteristics

Domain
laboratory
Configuration
benchtop
Display
digital, electronic
Other characteristics
heating, cooling, connected, circulating, WLAN, modular, remote maintenance, with natural refrigerant, with smartphone application connectivity, programmable
Temperature range

Max.: 200 °C
(392 °F)

Min.: -20 °C
(-4 °F)

Description

Product description
The Universa U 420 P bath circulation thermostat is supplied with a bath cover and pump connections as standard. A front drain valve located behind the ventilation grille simplifies safe replacement of the temperature control fluid.

Features
  • Bath circulation thermostat with advanced electronics for precise temperature control
  • Dedicated Tmax button for direct access to overtemperature protection settings during commissioning
  • Energy-efficient refrigeration: capillary tube technology with speed-controlled inverter compressor and speed-controlled fan
  • Includes programmer, calibration options and weekly scheduler for flexible process control
  • Integrated WLAN for connection to PC, tablet and smartphone
  • Modular design with interchangeable control head and heating/cooling base for future adaptation
  • Vario pump with six adjustable performance levels; flexible flow distribution for internal and external circulation
  • Use of natural, environmentally friendly refrigerants
  • Integrated web server for secure remote control and monitoring (PKI and 2‑factor authentication); connection to LAUDA.LIVE for cloud-based data analysis and remote maintenance
  • Comprehensive safety concept: float-level protection, adjustable overtemperature protection with acoustic alarm and automatic limit-value adjustment for tempering media
  • 3.5-inch colour TFT display with progression graph, six-language menu and intuitive cursor/soft-key operation
  • Front switch on the control head for pump output adjustment during operation
  • Adaptive bath-edge ventilation (patent pending) to prevent icing and condensation and to protect electronics
  • Flexible networking: integrated Ethernet and USB plus eleven optional interface modules; Wi‑Fi use limited to certified countries/regions


Working range
  • Min. working temperature: -20 °C
  • Max. working temperature: 200 °C
  • Temperature stability: 0.02 ± K


Cooling capacity table
Temperature | Pump level | Tempering medium | Cooling capacity 50 Hz | Cooling capacity 60 Hz
20 °C | 6 | Ethanol | 0.2 kW | 0.2 kW
10 °C | 6 | Ethanol | 0.19 kW | 0.19 kW
0 °C | 6 | Ethanol | 0.18 kW | 0.18 kW
-10 °C | 6 | Ethanol | 0.14 kW | 0.14 kW
-20 °C | 3 | Ethanol | 0.07 kW | 0.07 kW


Technical specifications
  • Working temperature range (DIN 12876): -20 … 200 °C
  • Ambient temperature range: 5 … 40 °C
  • Temperature stability: 0.02 ± K
  • Max. power consumption: 2.9 kW; Current consumption: 12 A
  • Bath opening (W×D): 130 x 100 mm; Bath volume min./max.: 1.8 / 4.0 L
  • Max. discharge pressure: 0.6 bar; Max. pump flow: 22 L/min
  • Inlet/outlet connection thread (outside): M16 x 1
  • Dimensions (W×D×H): 210 x 410 x 616 mm; Weight: 26 kg
  • Power supply: 220-240 V; 50/60 Hz; example power cord: angled plug (CEE7/7)