The Silbermann Jet Scavenging Inlet helps safely remove excess anesthetic gases from the operating theater, supporting a cleaner and safer environment for medical staff.
This inlet is designed to meet the latest, strict NFPA 99 safety requirements.
The scavenging inlet assembly uses compressed air to generate suction through the Venturi effect, drawing contaminated air away and discharging it safely to the outside atmosphere.
The jet scavenging inlet can be supplied recessed, panel-mounted, or exposed, with or without hose assembly.
FEATURES & BENEFITS- Operation based on compressed air - No need for a central AGSS system.
- When the probe is connected, a green pneumatic indicator shows the system is in use.
- Very quiet operation.
- Automatic shut-off when the probe is disconnected.
- Flow and vacuum adjustment.
- The locking mechanism quick connect.
TECHNICAL DATA- Inlet pressure: 3.5-5 bar
- Inlet flow: 40 LPM
- Free air suction capacity: Adjustable 35 -80 LPM
- Vacuum level: Adjustable 20-50 cmH2O
- Inlet pipe diameter: OD 3/8”
- Exhaust pipe diameter: OD 7/8”
Material of construction:- Box: Baked-paint steel
- Front Panel: Anodized Aluminum
- Injector: Brass
MEDICAL GAS SYSTEM STANDARDS:NFPA 99, ISO 7396, G-01, HTM 02-01
SILBERMANN’S QUALITY MANAGEMENT SYSTEM:EN ISO 13485-2016, ISO 9001-2015, CE MARKED