Technical Specifications
Safety: IEC60601 approval (GB9706)
Dimension: 340mm(W) x270mm(H) x230mm(D) Power Supply: 100-240V AC50Hz/60Hz Alarm Indication Visual and audible alarm 1000 groups of alarm events review Application: Adult, pediatric and neonate
ECG
Leads I. II .lll.aVR. aVL.aVF.V
Gains: 2imm/mv. 5mm/mv,10mmVmv.20mm/mv. 40mm/mv. auto
Input impedance: ^5Mii
Waveform display: 1-7
Common Mode Rejection Ratio: ^100dB
Bandwith: o.5-40Hz(monitoring mode)
0 05- 120Hz(diagnostic mode)
1-25Hz (surgical mode)
Protection: Segregation breakdown voltage 4000VAC. 50/60Hz, Resistant to high-frequency electrotome, myoelectriclty. AC. drifting interference Indication of Detachment of Electrodes:Availab!e Playback of Cardiogram: Available Scanning Speed: 12.5mm/s, 25mm/s. 50mm/s
NIBP
Method: Automatic oscillometnc
Operation Modes: Single(manual).cycle(automatic), rapid (continuous) Measurement: Systolic /Diastolic / Mean
Rangé: Adult : 10-2/OmmHg.
Pediatric: 1 0-235 mmHg Neonate: 10-135 mmHg Unit: mmHg / Kpa selectable
Auto Measurement: 1,2,3.4.5.10,15.20.30.45.60.90 minutes, 2,4.8 hours 500 groups of NIBP measuring data review
RESP
Method: Impedance Measuring Range: 0-120 times/min
Alarm Limit: 0*120rpm. upper and lower limit continuously adjjstable Apnea alarm: Available