• Products
  • Catalogs
  • News & Trends
  • Exhibitions

Clean bench for the cosmetics industry KVM series
for tissue culturefor the pharmaceutical industryfor the food industry

Clean bench for the cosmetics industry - KVM series - AIRTECH JAPAN, Ltd. - for tissue culture / for the pharmaceutical industry / for the food industry
Clean bench for the cosmetics industry - KVM series - AIRTECH JAPAN, Ltd. - for tissue culture / for the pharmaceutical industry / for the food industry
Clean bench for the cosmetics industry - KVM series - AIRTECH JAPAN, Ltd. - for tissue culture / for the pharmaceutical industry / for the food industry - image - 2
Add to favorites
Compare this product

Characteristics

Applications
for the cosmetics industry, for tissue culture, for the pharmaceutical industry, for the food industry, for clean rooms, for hospitals, for research laboratories
Configuration
benchtop
Filter type
HEPA
Air flow type
laminar flow
Structure material
stainless steel
Other characteristics
UV, with sliding front sash
Face velocity

Max.: 0.4 m/s
(1.3 ft/s)

Min.: 0.3 m/s
(1.1 ft/s)

Power

60 W, 90 W
(0 hp, 0 hp)

Length

75 cm, 100 cm
(29.53 in, 39.37 in)

Description

Product application
Bench-top clean bench available in two widths: 750 mm (KVM-757) and 1000 mm (KVM-1007). Features a slanted sliding shutter for improved ergonomics and DC brushless motor for reduced power consumption.

Key points
  • Slanted sliding shutter design enhances operator access and ergonomics.
  • DC brushless motor reduces power consumption by more than 30% compared with conventional units.

Specifications (KVM-757)
Model: KVM-757
Cleanliness: ISO-5 (Class 100)
Collection efficiency: ≥99.99% for 0.3 µm particles
Filter elements: Main filter – low-pressure-loss HEPA; Pre-filter – round saran net filter
Airflow: approx. 5.0 m³/min
Air velocity: approx. 0.34 m/s
Power supply: AC100V 1Φ 50/60Hz
Power consumption: approx. 60 W (excluding outlet)
Blower: 1 unit (DC brushless motor)
LED lighting: 5.3 W × 2
Germicidal lamp: 15 W × 1 (interlocked to operate only when shutter is fully closed)
Circuit protector: 10 A (including outlet)
Weight: approx. 50 kg
Construction: Body – painted steel; Worktop – SUS304 (hairline finish)

Specifications (KVM-1007)
Model: KVM-1007
Cleanliness: ISO-5 (Class 100)
Collection efficiency: ≥99.99% for 0.3 µm particles
Filter elements: Main filter – low-pressure-loss HEPA; Pre-filter – round saran net filter
Airflow: approx. 8.0 m³/min
Air velocity: approx. 0.40 m/s
Power supply: AC100V 1Φ 50/60Hz
Power consumption: approx. 90 W (excluding outlet)
Blower: 2 units (DC brushless motors)
LED lighting: 8.2 W × 2
Germicidal lamp: 15 W × 1 (interlocked to operate only when shutter is fully closed)
Circuit protector: 15 A (including outlet)
Weight: approx. 60 kg
Construction: Body – painted steel; Worktop – SUS304 (hairline finish)

Features / Technical specifications
  • Series: KVM series (bench-top clean benches: KVM-757, KVM-1007)
  • Applications: electronics/semiconductors & LCD, pharmaceuticals & cosmetics, hospitals, food, universities & research institutes, regenerative medicine and other clean work environments
  • Performance: ISO-5 (Class 100) level cleanliness; particle collection ≥99.99% at 0.3 µm
  • Energy saving: DC brushless motors for significantly reduced power consumption
  • Safety: germicidal lamp interlocked to operate only when shutter is fully closed
  • Materials: painted steel body with SUS304 worktop (hairline finish)
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.