Short descriptionBiological indicator ampoules with a very short incubation period. Ampoules contain bacterial spores. After sterilization and a 10-hour incubation, spore survival is assessed.
Detailed descriptionThe ampoules in the test pack contain bacterial strains and culture medium. The packaged biological indicator must be placed in the sterilizer chamber. After the sterilization cycle and addition of the medium, the ampoules are incubated at 55–60°C. Preliminary results can be read after 2.5 hours of incubation, with a definitive assessment of microbial growth after 10 hours. Preservation of the medium's original color in the biological ampoule confirms inactivation of microorganisms, i.e. effective sterilization.
Advantages- reliable test result, suitable for environments with strict infection-control requirements
- short incubation period for the biological material (preliminary read at 2.5 h, final at 10 h)
Additional informationThe test packaging also includes a chemical indicator that confirms the execution of sterilization.
Packing / variants (example gross prices shown on the website)50 units | 849,00 zł
10 units | 173,91 zł
1 unit | 19,99 zł
Technical characteristics / specifications- sterilization method to be verified: steam
- compliance with standards: ISO 11138, EN 867-3, ANSI/AAMI/ISO 14161:2000
- incubation temperature: 55–60°C
- read times: preliminary after 2.5 h, final after 10 h
- storage: ≤ 27°C
- quantity per pack: 50 units, 10 units or 1 unit
- package includes a chemical indicator confirming sterilization