Product overviewDental Class B benchtop steam autoclave Labotemp LDA-02 is engineered for dental clinics, medical practices and diagnostic/research laboratories. Using saturated steam with triple pre‑vacuum cycles, it provides validated sterilization for dental handpieces, hollow and surgical instruments, cassettes, wrapped loads and porous materials. The compact benchtop format and intuitive controls support traceable sterilization workflows required in regulated clinical environments.
Design and operationThe unit integrates a corrosion‑resistant stainless steel chamber, a rapid built‑in steam generator and microcomputer control with an LCD touch interface. Built‑in BD, Helix and vacuum test functions, USB/printer/U‑disk output and automated water fill, air discharge and steam exhaust simplify validation and record keeping. Multiple safety interlocks (over‑temperature, over‑pressure, low‑water, door lock) and intelligent diagnostics protect operators and loads. Benchtop footprint, low power draw and 110V/220V (50Hz/60Hz) compatibility ease installation in clinics and labs.
Key features- Safety door lock
- Corrosion‑resistant stainless steel inner chamber with sealed cover
- Advanced microcomputer control system
- LCD display with responsive touch operation
- European Class B sterilization standard
- Triple pre‑vacuum system for efficient air removal
- BD test, Helix test and vacuum test functions
- Built‑in steam generator for rapid saturated steam production
- Automatic water filling system
- Automatic cold air discharge and steam exhaust after sterilization
- Integrated standard test interface
- USB, U‑disk and printer support for sterilization data recording
- Multiple safety protections: over‑temperature, over‑pressure, low water level and anti‑dry burning protection
Applications- Dental clinics
Validated sterilization of dental handpieces, hollow instruments, wrapped surgical tools, cassettes and porous materials. - Hospitals & medical practices
Reliable sterilization for compatible reusable medical instruments and equipment. - Diagnostic laboratories
Sterilization of laboratory glassware, instruments and other materials requiring Class B steam sterilization. - Research laboratories
Supports validated sterilization workflows for laboratory equipment and reusable items. - Veterinary clinics
Sterilization of compatible reusable veterinary instruments and equipment. - Clinical & healthcare facilities
Designed for workflows needing pre‑vacuum air removal, process validation and traceable sterilization records.
Technical specifications- Capacity: 12 L
- Chamber volume: Ø200 × 355 mm
- Rated working temperature: 134°C
- Rated working pressure: 0.22 MPa
- Design temperature: 139°C
- Design pressure: 0.25 MPa
- Voltage: 110V/220V, 50Hz/60Hz
- Power: 2 kW
- Outer dimension: 445 × 580 × 400 mm
- Packing size (L × W × H): 650 × 500 × 450 mm
- Net weight: 44 kg