Product descriptionThe proven specialist for reliable ICH stability tests in the pharmaceutical and laboratory sector: the KBF 260 performs long‑term tests, accelerated tests and stress tests with precise climate control and comprehensive documentation options. Suitable for all installation locations thanks to flexible water supply variants. The latest KBF consumes up to 40 % less energy than the previous model. Optional light modules convert the unit for ICH photostability testing or plant growth applications.
Benefits- Suitable for critical ICH stability testing due to maximum climate homogeneity, even when fully loaded
- Reliable long‑term operation ensured by proven compressor technology and a stainless‑steel inner chamber
- User‑friendly: illuminated status indicator and ergonomic touchscreen program controller
- Energy‑efficient: up to 40% lower energy consumption versus the predecessor through inverter cooling technology
Important characteristics- Temperature range: 0 °C to +70 °C
- Humidity range: 10% to 80% RH
- APT.line™ preheating chamber technology
- Up to 40% lower energy consumption compared to the previous model
- Inverter cooling unit with climate‑neutral R‑600a refrigerant
- Door heating to prevent condensation
- Adjustable fan speed
- Humidity regulation with capacitive humidity sensor and responsive steam humidification
- Robust housing with HIT sandwich technology
- Stainless‑steel inner chamber and metal oxide coated safety glass inner door
- Two stainless‑steel racks; accessory shelves up to 9 positions
- 80 mm × 45 mm access port with silicone plug (other sizes available)
- Floor units with sturdy twin casters and front brakes
- Intuitive touchscreen controller with time‑segment and real‑time programming
- Internal data logger; measured values exportable via USB in open format
- Class 3.3 independent temperature safety device (DIN 12880) with visual and acoustic alarms
- Unit self‑test for comprehensive status analysis
- Data interfaces: Ethernet, USB; BINDER APT‑COM™ multi‑management software (basic edition)
- Optional ICH light modules and LED plant light modules
Performance data (selected)Article numbers: 9020‑0479 (standard) | 9020‑0495 (alternative voltage). Temperature range (basic): 0...70 °C. Max. heat compensation at 40 °C: 410 W. Temperature range with humidity: 10...70 °C. Humidity range: 10...80 % RH. Temperature uniformity at 25 °C / 60% RH: 0.3 ± K. Temperature uniformity at 40 °C / 75% RH: 0.2 ± K. Temperature fluctuation at 25 °C / 60% RH: 0.1 ± K. Temperature fluctuation at 40 °C / 75% RH: 0.1 ± K. Humidity fluctuation at 25 °C / 60% RH: 1.5 ± % RH. Humidity fluctuation at 40 °C / 75% RH: 1.5 ± % RH. Recovery time after door opened 30 s at 25 °C / 60% RH: ~2 min. Recovery time after door opened 30 s at 40 °C / 75% RH: ~3 min.
Electrical dataRated voltage: 220...240 V (9020‑0479) | 120 V (9020‑0495). Power frequency: 50/60 Hz. Nominal power: 1.4 kW. Unit fuse: 16.0 A. Phase: 1~.
Dimensions & weightsInterior volume: 262 L. Net weight (empty): 188 kg. Load per rack: 40 kg. Permitted load: 200 kg. Wall clearance back: 100 mm. Wall clearance sidewise: 180 mm. Width net: 930 mm. Height net: 1,396 mm. Depth net: 900 mm. Interior width × height × depth: 656 × 695 × 575 mm. Inner doors: 1. Unit doors: 1.
FixturesNumber of shelves (std./max.): 2 / 9.
Environment-specific dataSound‑pressure level at 25 °C / 60% RH: 50 dB(A). Energy consumption at 25 °C / 60% RH: 215 Wh/h (9020‑0479) | 255 Wh/h (9020‑0495). Energy consumption at 40 °C / 75% RH: 230 Wh/h (9020‑0479) | 295 Wh/h (9020‑0495).
Options & accessories (examples)- Reinforced inner chamber (permitted load with U‑rails 300 kg)
- Door lock (mechanical or electromechanical with PIN via controller)
- Interior lighting with climate‑resistant LEDs
- Swivel or height‑adjustable casters
- Shipping pallet with unloading ramp
- Access ports with silicone plugs (30 / 50 / 100 mm; left / right / top)
- Analog output 4‑20 mA (temperature/humidity)
- Specimen temperature display (flexible Pt100)
- Calibration and qualification services (various certificates and measurement plans)
- ICH photostability light modules, LED plant light modules, LED light bars
- CO₂ gassing/control modules and related installation services
Technical specifications- Model name: KBF 260
- Article numbers (examples): 9020‑0479 (230 V) / 9020‑0495 (120 V)
- Temperature range: 0...70 °C
- Humidity range: 10...80 % RH
- Inner usable volume: 262 L
- Inner dimensions (W×H×D): 656 × 695 × 575 mm
- Net dimensions (W×H×D): 930 × 1,396 × 900 mm
- Net weight (empty): 188 kg
- Power / fuse: 1.4 kW / 16 A
- Rated voltage: 220...240 V (std.) or 120 V (option)
- Data interfaces: Ethernet, USB; internal data logger with open‑format USB export
- Safety: Class 3.3 independent temperature safety device (DIN 12880) with visual and acoustic alarms
- Doors: 1 outer door, 1 inner door; inner door safety glass coated with metal oxide
- Shelves: 2 standard (up to 9 max); load per rack 40 kg; permitted load 200 kg
- Refrigerant: R‑600a (climate‑neutral)
- Energy‑efficient inverter cooling; up to 40% lower energy consumption vs predecessor
- Optional modules: ICH photostability light modules, LED plant light modules, CO₂ modules, APT‑COM™ multi‑management software
- Recovery times after 30 s door opening: ~2–3 minutes depending on conditions