Benefits- A reliable option for critical ICH stability testing thanks to maximum climate homogeneity, even when the unit is fully loaded
- Reliable: failsafe operation over the long term thanks to proven compressor technology and stainless-steel interior
- Smart: signal triangle with light-up status display and ergonomic control terminal with touchscreen program controller
- Economical: up to 40% lower energy consumption than the previous model thanks to modern inverter cooling technology
Product description The proven specialist for reliable ICH stability tests in the pharmaceutical industry: the KBF masters all long-term tests, accelerated tests and stress tests. It fulfills all important ICH requirements, from precision to programming and documentation. With flexible water supply options for all installation locations. In addition, the new KBF consumes up to 40 % less energy than its predecessor. With optional light modules, the KBF can be converted into a plant growth chamber.
Important characteristics- Temperature range: 0 °C to +70 °C
- Humidity range: 10% to 80% RH
- Proven APT.line™ preheating chamber technology
- Energy-efficient cooling with modern inverter cooling unit
- Cooling with climate-neutral R-600a refrigerant
- Door heating to prevent condensation
- Adjustable fan speed
- Humidity regulation with capacitive humidity sensor and responsive steam humidification
- Sturdy housing design with HIT sandwich technology
- Stainless-steel inner chamber
- Inner doors made of metal oxide-coated safety glass
- Two stainless-steel racks (standard)
- 80 mm x 45 mm access port with silicone plug
- Floor-standing units with sturdy twin casters with brakes at the front (standard for some configurations)
- Intuitive touchscreen controller with time-segment and real-time programming
- Ergonomically adjustable operating terminal
- Signal triangle with light-up status display
- Internal data logger; measured values can be read out in open format via USB
- Class 3.3 independent temperature safety device (DIN 12880) with visual and acoustic alarms
- Unit self-test for comprehensive status analysis
- Data interfaces: Ethernet, USB
- BINDER Multi Management Software APT-COM™ basic edition (available)
- Optional light modules for ICH photostability tests
- Optional LED light modules for plant growth
Technical data / specifications- Data (model variants): Article Number: 9020-0483 / 9020-0499
Designation: KBF1600-230V (9020-0483) / KBF1600UL-120V (9020-0499)
Option model: Standard - Performance Data Temperature: Temperature range: 0...70 °C
Max. heat compensation at 40 °C: 640 W - Performance Data Climate: Temperature range with humidity: 10...70 °C
Humidity range: 10...80 % RH
Temperature uniformity at 25 °C and 60% RH: 0.4 ± K
Temperature uniformity at 40 °C and 75% RH: 0.3 ± K
Temperature fluctuation at 25 °C and 60% RH: 0.1 ± K
Temperature fluctuation at 40 °C and 75% RH: 0.1 ± K
Humidity fluctuation at 25 °C and 60% RH: 1.5 ± % RH
Humidity fluctuation at 40 °C and 75% RH: 1.5 ± % RH
Recovery time after door opened 30 s at 25 °C / 60% RH: 3 min
Recovery time after door opened 30 s at 40 °C / 75% RH: 7 min - Electrical data: Rated Voltage: 220...240 V (9020-0483) / 120 V (9020-0499)
Power frequency: 50/60 Hz
Nominal power: 1.7 kW
Unit fuse: 16.0 A
Phase (Nominal voltage): 1~ - Dimensions and weights: Interior volume: 1,610 L
Net weight of the unit (empty): 438 kg
Load per rack: 40 kg
Permitted load: 200 kg
Wall clearance back: 100 mm
Wall clearance sidewise: 180 mm - Housing dimensions (net): Width net: 1,795 mm
Height net: 1,951 mm
Depth net: 1,175 mm - Internal Dimensions: Interior width: 1,521 mm
Interior height: 1,250 mm
Interior depth: 847 mm
Inner doors: 2
Unit doors: 2 - Environment-specific data: Sound-pressure level at 25 °C and 60% RH: 54 dB(A)
Energy consumption at 25 °C and 60% RH: 395 Wh/h (for 9020-0483) / 410 Wh/h (other variant)
Energy consumption at 40 °C and 75% RH: 460 Wh/h (9020-0483) / 570 Wh/h (9020-0499) - Fixtures: Number of shelves (std./max.): 2 / 16
- Accessories & options (selection): Door lock (8012-2536); Electromechanical door lock (8012-2589); Interior lighting (8012-2569); Swivel casters, height-adjustable (8012-2543); Shipping pallet with ramp (8012-2547); Access ports various sizes/positions (e.g., 8012-2483, 8012-2486, 8012-2488); Inner chamber, reinforced (8012-2423); Analog output 4-20 mA (8012-2574); Specimen temperature display (8012-2591); Calibration and certification services (multiple items)