Benefits- Safe: tested according to ICH Q1A guideline, including full-load scenarios.
- Reliable: failsafe design for continuous operation and status self-check.
- Adaptable: wide range of accessories and configurable racks for customer-specific setups.
- Low energy consumption: optimized thermoelectric (Peltier) cooling reduces power use and noise.
Product descriptionThe KBF-S ECO 400 constant climate chamber uses thermoelectric (Peltier) cooling with patented heat dissipation to provide precise temperature and humidity control in a compact 400 L cabinet. The design yields low operating noise and minimal energy draw while offering features tailored for laboratory and industrial stability testing.
Important characteristics- Temperature range: 0 °C to +70 °C (max. 26 °C under ambient temperature)
- Humidity range: 10% to 80% RH
- APT.line™ preheating chamber technology
- Independent water supply via tank
- LCD displaying temperature, humidity, alarms and additional information
- Internal data logger; measured values exportable in open format via USB
- Unit self-test for comprehensive status analysis
- Inner chamber completely stainless steel; 2 stainless steel racks
- Access port with silicone plug Ø 30 mm, left
- Computer interface: Ethernet
Technical specifications- Article Number: 9020-0463 (designation KBFSECO400-230V) / 9020-0464 (designation KBFSECO400UL-120V)
- Designation / Model codes: KBFSECO400-230V (230 V) ; KBFSECO400UL-120V (100...120 V)
- Option model: Standard
- Temperature range (performance): 0...70 °C (max. 26 °C under ambient temperature)
- Temperature range with humidity: 5...70 °C (max. 26 °C under ambient temperature)
- Humidity range: 10...80 % RH
- Max. heat compensation at 40 °C: 300 W
- Temperature uniformity (25 °C, 60% RH): 0.1 ± K
- Temperature fluctuation (25 °C, 60% RH): 0.1 ± K
- Humidity fluctuation (25 °C, 60% RH): 0.3 ± % RH
- Recovery time after door open 30 s (25 °C, 60% RH): 6 min
- Rated voltage: 200...230 V (9020-0463) / 100...120 V (9020-0464)
- Power frequency: 50/60 Hz
- Nominal power: 1.1 kW
- Unit fuse: 16.0 A
- Phase (Nominal voltage): Single-phase (1~)
- Interior volume: 400 L
- Net weight (unit empty): 192 kg
- Load per rack: 30 kg
- Permitted load (total): 120 kg
- Wall clearance back: 100 mm
- Wall clearance sidewise: 180 mm
- Housing width (net): 925 mm
- Housing height (net): 1,946 mm
- Housing depth (net): 796 mm
- Interior width: 650 mm
- Interior height: 1,268 mm
- Interior depth: 485 mm
- Unit doors: 1
- Number of shelves (std./max.): 2 / 15
- Sound-pressure level: 45 dB(A)
- Energy consumption (25 °C, 60% RH): 60 Wh/h
- Energy consumption (40 °C, 75% RH): 70 Wh/h
- Other included features: APT.line™ preheating chamber technology; LCD display with alarms; internal data logger (USB export in open format); unit self-test; stainless steel inner chamber.