Product OverviewThe Fiv Felv Heartworm Fcov rapid test is a lateral flow immunochromatographic assay for the qualitative detection of feline immunodeficiency virus antibody (FIV Ab), feline leukemia virus antigen (FeLV Ag), feline Coronavirus antibody (FCoV Ab) and Heartworm antigen (CHW Ag) in cat's serum, plasma and whole blood specimen. Assay Time: 5-10 minutes.
Product DetailsProduct Name: Fiv Felv Heartworm Fcov rapid test
Specimen: cat's serum, plasma and whole blood specimen
Assay Time: 5-10 minutes
MOQ: 800 test kits
Delivery time: 1 week after Get payment
Packing: 10 test kits/Packing box
Storage Temperature: 2-30°C, DO NOT FREEZE. Do not store the test kit in direct sunlight.
Shelf Life: 18 months
Benefits- One-step whole blood, serum and plasma testing.
- Simple operation, non-professional on-site inspection.
- The detection process is fast, and the result will be displayed in about 15 minutes.
- The test time is less than or equal to 10 seconds, and it can detect 270 samples per hour at the fastest.
- The detection equipment is compact, easy to carry, mobile terminal operation, cloud platform quality control.
- The detection sensitivity is high, comparable to that of enzyme-labeled kits.
Components- Test devices
- Droppers
- Single Buffer
- IFU (Instructions for Use)
PrincipleThe rapid test is based on sandwich method lateral flow immunochromatographic assay. The test card has a testing window for observation of assay running and result reading with invisible T (test) zone(s) and a C (control) zone. When sample is applied, liquid will laterally flow through each test strip and react with pre-coated FIV recombinant antigens, FeLV antibodies, FCoV recombinant antigens and Heartworm antibodies respectively. If target analytes are present, a visible T line appears in the corresponding window. The C line should always appear after sample application to indicate a valid result.
Storage and StabilityThe kit can be stored at room temperature (2-30°C). The test kit remains stable through the expiration date (18 months) marked on the package label. DO NOT FREEZE. Do not store the test kit in direct sunlight.
Test Procedure- Allow all materials, including specimen and test device, to recover to 15-25°C before running the assay.
- Take out the test card from the foil pouch and place it horizontally.
- For FIV/FeLV/CHW: Collect 50 μL of the prepared specimen into a vial of FIV-FeLV-CHW assay buffer and mix well. Then drop 3 drops (approx. 120 μL) of the diluted sample into each sample hole "S" of the test card, corresponding to the windows FIV, FeLV, CHW. Start the timer.
- Interpret the result in 5-10 minutes. Result after 10 minutes is considered invalid.
- For FCoV Ab: Collect 20 μL of the prepared specimen into a vial of FCoV Ab assay buffer and mix well. Then drop 3 drops (approx. 120 μL) of the diluted sample into the sample hole "S" corresponding to the FCoV window. Start the timer.
- Interpret the result in 5-10 minutes. Result after 10 minutes is considered invalid.
Interpretation of Assay Result- Positive (+): Both the C line and the T line are present (T line may be clear or fuzzy).
- Negative (-): Only the sharp C line appears; no T line.
- Invalid: No color line in zone C, regardless of T line presence.
Clinical Data (summary)FIV (comparison vs Western Blot): Relative Sensitivity: 94.9%
Relative Specificity: 97.4%
Relative Accuracy: 96.6%
FeLV (comparison vs Western Blot): Relative Sensitivity: 96.7%
Relative Specificity: 99.4%
Relative Accuracy: 98.9%
FCoV Ab (comparison vs ELISA): Relative Sensitivity: 99.1%
Relative Specificity: 89.8%
Relative Accuracy: 92.9%
Heartworm Ag (comparison vs PCR): Relative Sensitivity: 92.4%
Relative Specificity: 98.0%
Relative Accuracy: 95.8%
Co mponents / Packaging (additional)• Test devices
• Droppers
• Single Buffer(s)
• IFU (Instructions for Use)
Caractéristiques / spécifications techniques- Denomination commerciale: Fiv Felv Heartworm Fcov Rapid Test
- Format: Lateral flow immunochromatographic assay
- Analyses détectées: FIV antibody (FIV Ab), FeLV antigen (FeLV Ag), FCoV antibody (FCoV Ab), Heartworm antigen (CHW Ag)
- Specimen: Cat's serum, plasma and whole blood
- Assay time / Result read: 5-10 minutes (results after 10 minutes considered invalid)
- Sample volumes: 50 μL for FIV/FeLV/CHW buffer preparation; 20 μL for FCoV Ab buffer preparation; then ~3 drops (~120 μL) into test well
- Packing: 10 test kits per packing box
- MOQ: 800 test kits
- Storage temperature: 2-30°C; do not freeze; protect from direct sunlight
- Shelf life: 18 months
- Throughput: up to 270 samples per hour at fastest (test time ≤ 10 seconds claimed)
- Performance: sensitivity and specificity metrics provided vs reference methods (Western Blot, ELISA, PCR)