Product descriptionA BINDER MKF 720 environmental simulation chamber for standardized temperature and climate testing according to DIN and IEC. Designed for rapid temperature changes with integrated humidity control and comprehensive standard equipment to support laboratory and production testing workflows.
Benefits- Uniform climate distribution via APT.line™ preheating chamber technology
- Automated fresh‑water and wastewater management
- Fast‑response steam humidification with capacitive humidity sensing
- Extensive programming, data logging and Ethernet connectivity (APT-COM™ compatible)
- Heated viewing window with LED interior lighting and user‑friendly touchscreen controller
Important characteristics- Temperature range −40 °C to +180 °C (with humidity: 10…95 °C)
- Humidity control range 10 % to 98 % r.H.
- Integrated 20 L water storage tank and automatic water management
- APT.line™ preheating chamber technology and programmable condensation protection
- 2 access ports (80 mm) with silicone plugs; heated viewing window
- Internal stainless steel chamber and 1 stainless steel rack (std.)
- CFC‑free refrigerant R‑452A and compressor cooling
Technical specifications- Article number (standard): 9020-0381; (with voltage/frequency converter): 9020-0359
- Designation: MKF720-400V (std.) / MKF720-480V-C (with converter)
- Temperature uniformity: 0.1…1.8 ± K (depending on set value)
- Temperature fluctuation: 0.1…0.5 ± K (depending on set value)
- Heating/cooling rate (IEC 60068-3-5) approx. 4.8 K/min (−40 °C ↔ 180 °C); cooling time 180 °C → −40 °C: 120 min
- Humidity fluctuation ≤2.5 ± % r.H.; temperature fluctuation with humidity 0.2…1.5 ± K
- Rated voltages: 400 V (50 Hz) standard; 480 V (60 Hz) option; nominal power 11 kW; device fuse 32 A
- Interior volume: 734 L; interior dimensions (W×H×D): 1,200 × 1,020 × 600 mm
- Net weight (empty): 590 kg; permitted total load: 160 kg; load per rack: 40 kg
- Number of shelves (std./max.): 1 / 11; 2 access ports 80 mm (left/right)
- Sound pressure level: 65 dB(A) / 67 dB(A) (option); energy consumption at 25 °C / 60 % RH: 3,000 Wh/h
- Interfaces: Ethernet; internal data logger with USB export (open format); intuitive touchscreen controller; APT-COM™ 4 Basic included