Product descriptionThe BINDER MK 720 CO₂ climate test chamber performs cyclical heat and cold tests over a wide temperature range. It is designed for rapid temperature changes and for reproducible environmental conditioning of samples in industrial and laboratory use.
Benefits- Safe: reliable protection for samples, data and processes
- Reliable: durable components, norm-compliant performance data and long-term functional stability
- Smart: intuitive touchscreen operation, digital networking and modular equipment
- Economical & ecological: energy-efficient CO₂ (R-744) cooling technology and cost-effective operation
Key features- Temperature range: -48 °C to +180 °C
- APT.line™ preheating chamber technology
- Cooling with climate-neutral refrigerant R-744 (CO₂)
- Intuitive touchscreen controller with time-segment and real-time programming
- Ergonomically adjustable operating terminal and signal triangle with illuminated status display
- Internal data logger; measured values exportable in open format via USB; Ethernet and USB interfaces
- BINDER Multi Management Software APT-COM™ (basic edition)
- Programmable condensation protection and heated viewing window with LED interior lighting
- Stainless-steel inner chamber and 1 stainless-steel rack (std.), up to 11 shelves
- Access ports with silicone plugs: 2 × 80 mm (left and right) for model 720
- Class 2 independent adjustable temperature safety device (DIN 12880) with visual alarm; troubleshooting system with visual and acoustic alarms
- 4 sturdy twin casters with brakes and door heating to prevent condensation
Technical specifications- Article Number: 9020-0522 / 9020-0524
- Designation: MK720CO2-400V
- Version: Standard
- Temperature range: -48...180 °C
- Temperature uniformity (dependent on set value): 0.1...1.5 ± K
- Temperature fluctuation (dependent on set value): 0.1...0.5 ± K
- Average heating rate (IEC 60068-3-5) -40→180 °C: 4.5 K/min
- Average cooling rate (IEC 60068-3-5) 180→-40 °C: 5.0 K/min
- Cooling time 180→-40 °C: 90 min
- Rated voltage: 400 V (9020-0522) / 480 V (9020-0524)
- Power frequency: 50 Hz
- Nominal power: 8.7 kW
- Unit fuse: 16.0 A
- Phase (nominal voltage): 3~/N/PE
- Interior volume: 734 L
- Net weight (empty): 495 kg
- Load per rack: 40 kg; permitted total load: 160 kg
- Wall clearance back/side: 300 mm / 200 mm
- Width × Height × Depth (net): 1,577 × 2,005 × 1,133 mm
- Interior (W×H×D): 1,200 × 1,019 × 600 mm
- Unit doors: 1; Shelves (std./max.): 1 / 11
- Sound-pressure level: 65 dB(A)
- Energy consumption at 25 °C: 2,300 Wh/h