Compare this product

Characteristics

Type
climatic
Controlled parameter
CO2
Test type
stress
Applications
for materials testing
Configuration
floor-standing, mobile
Number of doors
1-door
Chamber material
stainless steel
Other characteristics
high-performance
Capacity

228 l
(60.2 gal)

Temperature range

Max.: 180 °C
(356 °F)

Min.: -40 °C
(-40 °F)

Description

Product description
The BINDER MK series chamber is designed for heat and cold tests between -48 °C and +180 °C. The MK 240 CO₂ provides a compact, standardised solution for temperature cycling and rapid temperature changes, combining APT.line™ preheating technology with climate-neutral R-744 (CO₂) cooling.

Benefits
  • Safe: reliable protection for samples, data and processes
  • Reliable: durable components, standard-compliant performance and lasting functional stability
  • Smart: intuitive touchscreen control, digital networking and configurable equipment
  • Economical & ecological: competitive price/performance, energy-efficient CO₂ refrigeration and future-proof options


Key features
  • Operational temperature range: -48 °C to +180 °C (technical range -50...180 °C)
  • APT.line™ preheating chamber technology
  • Cooling with climate-neutral refrigerant R-744 (CO₂)
  • Intuitive touchscreen controller with time-segment and real-time programming
  • Ergonomically adjustable operating terminal and signal triangle with illuminated status display
  • Internal data logger; measured values exportable in open format via USB
  • BINDER Multi Management Software APT-COM™ basic edition
  • Programmable condensation protection and door heating to avoid condensation
  • Stainless-steel inner chamber and heated viewing window with LED interior lighting
  • Access port(s) (models 240, 400): 50 mm left; additional port options available
  • Class 2 independent adjustable temperature safety device (DIN 12880) with visual alarm
  • 4 twin casters with front brakes
  • Data interfaces: Ethernet, USB; optional RS-485
  • Adjustable ramp function; integrated line recorder; real-time clock


Data
Article Number: 9020-0513 / 9020-0515
Designation: MK240CO2-400V
Option model: Standard

Performance Data — Temperature
Temperature range: -50...180 °C
Temperature uniformity (depending on set value): 0.1...1.5 ± K
Temperature fluctuation (depending on set value): 0.1...0.5 ± K
Average heating rate (IEC 60068-3-5) between -40 °C and 180 °C: 5.0 K/min
Average cooling rate (IEC 60068-3-5) between 180 °C and -40 °C: 5.0 K/min
Cooling time 180 °C → -40 °C: 90 min

Electrical data
Rated voltage: 400 V (variant: 480 V)
Power frequency: 50 Hz
Nominal power: 5.2 kW
Unit fuse: 16.0 A
Phase (nominal voltage): 3~/N/PE

Dimensions and weights
Interior volume: 228 L
Net weight (empty): 311 kg
Load per rack: 30 kg
Permitted total load: 70 kg
Wall clearance back: 300 mm
Wall clearance sideways: 200 mm

Housing dimensions (not incl. fittings and connections)
Width net: 1,112 mm
Height net: 1,713 mm
Depth net: 926 mm

Internal Dimensions
Interior width: 735 mm
Interior height: 700 mm
Interior depth: 442 mm
Unit doors: 1

Environment-specific data
Sound-pressure level: 62 dB(A)
Energy consumption at 25 °C: 1,000 Wh/h

Fixtures
Number of shelves (std./max.): 1/6

Options, accessories & services (selection)
  • Analog output 4-20 mA for temperature values (8012-2634)
  • Access ports with silicone plugs — various diameters and positions (e.g. 30/50/80/100/125 mm)
  • Additional Pt100 temperature sensor (8012-2667)
  • Door lock (8012-2669)
  • Calibration certificates and temperature measurement packages; maintenance contracts (BRONZE/SILVER/GOLD)
  • Commissioning, IQ/OQ/PQ and leak test services


Technical specifications — summary
  • Model: MK 240 CO₂
  • Operational temperature range: -48 °C to +180 °C (technical range -50...180 °C)
  • Interior volume: 228 L
  • Internal dimensions (W×H×D): 735 × 700 × 442 mm
  • Net weight (empty): 311 kg
  • Rated voltage: 400 V (option: 480 V)
  • Nominal power: 5.2 kW
  • Refrigerant: R-744 (CO₂)
  • Heating/cooling rate (IEC 60068-3-5): approx. 5.0 K/min
  • Interfaces: Ethernet, USB (optional RS-485)
  • Standard shelves: 1 (up to 6 max)

Catalogs

Exhibitions

Meet this supplier at the following exhibition(s):

Lab-Supply Dresden
Lab-Supply Dresden

8-08 Sep 2026 Dresden (Germany)

  • More information
    Lab-Supply Hamburg
    Lab-Supply Hamburg

    3-03 Nov 2026 Hamburg (Germany)

  • More information
    *Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.