Product descriptionThe BD Avantgarde.Line BD 56 is a bacteriological incubator engineered for gentle incubation of microorganisms and conditioning of heat‑sensitive media. Compact (57 L) and stackable, it delivers homogeneous temperature distribution suitable for laboratory and clinical workflows.
Benefits- High temporal and spatial temperature accuracy for reproducible results
- Reliable processes with integrated safety device and visual alarm
- 100 °C disinfection routine for in‑place sanitation
Key features- Temperature range: +5 °C above ambient to +100 °C
- APT.line™ control technology for enhanced temperature accuracy
- Natural convection
- Controller with LCD display and USB data port
- Electromechanical exhaust air flap control
- Inner door in tempered safety glass (ESG)
- 2 chrome‑plated racks (standard), stackable design
- Integrated independent temperature safety device Class 3.1 (DIN 12880) with visual alarm
Technical specifications (key data)Article Number: 9010-0323 / 9010-0324
Designation: BD056-230V / BD056UL-120V
Model: Standard
Temperature range: +5 °C above ambient to 100 °C
Temperature uniformity at 37 °C: 0.3 ± K
Temperature uniformity at 100 °C: 1.1 ± K
Temperature fluctuation at 37 °C: 0.1 ± K
Temperature fluctuation at 100 °C: 0.3 ± K
Heating time to 100 °C: 75 min
Heating time to 37 °C: 52 min
Recovery time after 30 s door opening at 37 °C: 15 min
Rated voltage: 230 V (9010-0323) / 120 V (9010-0324)
Power frequency: 50/60 Hz (230 V) / 60 Hz (120 V)
Nominal power: 0.3 kW
Unit fuse: 6.3 A (230 V) / 12.5 A (120 V)
Phase: 1~
Interior volume: 57 L
Net weight (empty): 39 kg
Load per rack: 30 kg
Permitted total load: 70 kg
Wall clearance back: 160 mm
Wall clearance sidewise: 100 mm
Housing (W×H×D) net: 560 × 625 × 565 mm
Internal (W×H×D): 360 × 420 × 380 mm
Inner doors: 1
Unit doors: 1
Number of shelves (std./max.): 2 / 4
Energy consumption at 100 °C: 110 Wh/h
Energy consumption at 37 °C: 25 Wh/h
Options & accessories- Access ports with silicone plugs (various positions/diameters)
- Calibration certificates (multiple point options)
- Additional PT100 sensor
- 4–20 mA analog temperature output
- Overtemperature alarm function
- Ethernet interface for APT‑COM™
- Lock, transport trolley, stainless steel/perforated shelves
- APT‑COM™ software (BASIC/PROFESSIONAL/GLP), IQ/OQ/PQ documents, service contracts
Technical summary- Model: BD 56 — BD Avantgarde.Line
- Intended use: bacteriological incubation and conditioning of heat‑sensitive media
- Convection: natural
- Control: APT.line™ with LCD; USB data logging
- Safety: independent temperature safety device Class 3.1 (DIN 12880)
- Volume: 57 L; nominal power: 0.3 kW; voltage options: 230 V / 120 V