Product descriptionThe MKF 720 CO₂ climate test chamber is designed for combined dynamic temperature and humidity tests with high change rates and precise humidity control. It supports reproducible, standards‑compliant test sequences and is equipped with a comprehensive set of standard features. The refrigeration circuit uses natural CO₂ (R‑744) suitable for applications with specific cooling requirements.
Benefits- Optimised performance-to-cost ratio and long-term investment security
- Easy integration, intuitive operation and flexible networking
- Reliable protection for samples, data and processes
- Reproducible results and long-term operational stability
Key characteristics- Temperature range: -48 °C to +180 °C
- Humidity range: 10 % r.H. to 98 % r.H.
- Humidity control with capacitive sensor and steam humidification
- APT.line™ preheating chamber technology
- Cooling with climate-neutral R‑744 (CO₂) refrigerant
- Intuitive touchscreen controller with time-segment and real-time programming
- Ergonomically adjustable operating terminal
- Signal triangle with illuminated status display
- Internal data logger; measured values exportable via USB in open format
- BINDER Multi Management Software APT-COM™ Basic Edition
- Programmable condensation protection and heated door to prevent condensation
- Troubleshooting system with visual and acoustic alarms
- Stainless-steel interior
- 1 stainless-steel rack (std.); shelves (std./max.): 1 / 11
- 2 access ports with silicone plugs, 80 mm (left and right)
- Heated viewing window with LED interior lighting
- Class 2 independent adjustable temperature safety device (DIN 12880) with visual alarm
- 4 robust twin casters, front brakes
- Data interfaces: Ethernet, USB
- Adjustable ramp function; integrated line recorder; real-time clock
Technical specifications- Article numbers: 9020-0523 (400 V) / 9020-0525 (480 V)
- Designation: MKF720CO2-400V / MKF720CO2-480V
- Version: Standard
- Temperature range: -48...180 °C
- Temperature uniformity (dependent on set value): 0.1...1.0 ± K
- Temperature fluctuation (dependent on set value): 0.1...0.2 ± K
- Average heating rate (IEC 60068-3-5 between -40 °C and 180 °C): 5.1 K/min
- Average cooling rate (IEC 60068-3-5 between 180 °C and -40 °C): 5.7 K/min
- Cooling time from 180 °C to -40 °C: 60 min
- Rated voltage: 400 V / 480 V
- Power frequency: 50 Hz (400 V) / 60 Hz (480 V)
- Nominal power: 11 kW
- Unit fuse: 25.0 A
- Phase (Nominal voltage): 3~/N/PE
- Interior volume: 734 L
- Internal usable dimensions: W 1,200 mm; H 1,020 mm; D 600 mm
- Housing net dimensions (without fittings/connections): W 1,580 mm; H 2,005 mm; D 1,140 mm
- Net weight (empty): 518 kg
- Load per rack: 40 kg; Permitted total load: 160 kg
- Wall clearance (back): 300 mm; Wall clearance (side): 200 mm
- Sound-pressure level: 60 dB(A)
- Energy consumption at 25 °C and 60 % r.H.: 3,400 Wh/h
- Water reservoir: 20 L
- Optional accessories: reinforced stainless-steel racks, access ports (30/50/80/100/125 mm), door lock, analog outputs 4-20 mA, additional Pt100 probe, RS485 interface, condensate reuse water circuits, calibration certificates, APT-COM™ software editions