Product overview Fixed ducted fume cupboards are fixed in position and suitable for chemistry labs, preparation rooms and classrooms in schools and colleges where harmful fumes need to be extracted from the work area. They provide 100% exhaust to atmosphere and use durable filters that do not require annual replacement, resulting in low running costs and minimal consumable maintenance.
Compliance - CLEAPSS Guide G9 – Fume Cupboards in Schools (Revision of DfEE Building Bulletin 88)
- BS EN 14175
- BREEAM compliance (ENE11 & HEA17)
- CLEAPSS tested
- Good Practice Guide 320 – Energy Efficient Design and Operation of Fume Cupboards
Key features - Visibility to both sides suitable for demonstrations
- All fumes extracted to atmosphere (100% exhaust)
- Trespa Athlon bunded work surface
- Vertically sliding toughened glass sash
- Built-in storage compartment
- Integral high level lighting
- Airflow control panel
- RCD-protected double socket outlet
- Gas outlet
- Water outlet
- Drip cup
- Dimension: 1000 mm wide x 715 mm deep x 2070 mm high
- Nominal air flow: 0.4 m/s face velocity at a 400 mm opening
- Voltage: 240V / 1 ph / 50Hz
Technical specifications - Application: educational chemistry labs, prep rooms, classrooms
- Exhaust type: fixed ducted, 100% exhaust to atmosphere
- Work surface: Trespa Athlon (bunded)
- Sash: vertically sliding toughened glass
- Lighting: integral high level lighting
- Controls: airflow control panel
- Electrical: RCD protected double socket outlet
- Utilities: gas outlet, water outlet, drip cup
- Dimensions (W x D x H): 1000 mm x 715 mm x 2070 mm
- Nominal face velocity: 0.4 m/s at 400 mm opening
- Electrical supply: 240V / 1 ph / 50Hz