Overview For smaller schools and colleges where a fume cupboard may not be in constant use or space is at a premium, mobile recirculating fume cupboards from Envair Technology provide a compact, flexible solution. The units feature glass panels on all four sides for 360° visibility of experiments and are designed to pass comfortably through standard door openings for easy movement between labs or classrooms. When required for use the unit locks into a docking station; when lessons finish it can be returned against a wall. The robust lockable docking station houses water, gas, drainage and power connections with quick-release fittings to enable safe, repeatable connection and disconnection. The hood recirculates laboratory air via an integral fan and a 25 kg carbon filter suitable for standard GCSE and A Level chemicals. Annual maintenance including carbon filter saturation testing and replacement is available in line with COSHH requirements. All Envair Technology educational fume cupboards are designed to CLEAPSS Guide G9 and are independently tested and approved by CLEAPSS.
Key Features - Fully mobile with locking castors for secure positioning and easy transport between classrooms
- Glass panels on all four sides for 360° visibility—ideal for demonstrations
- Integral 25 kg carbon filter with protective pre-filter
- Vertically sliding toughened glass sash for operator protection
- Integral high-level lighting for clear demonstration visibility
- Airflow control panel for monitoring and adjustment
- RCD-protected double socket outlet
- Remote-operated gas and water outlets
- Drip cup, bottle trap and flexible drain hose included
- Stainless steel restraining tether for secure installation
- Dimensions: 1000 mm wide × 715 mm deep × 1950 mm high (also available in 1200 mm and 1500 mm widths)
- Nominal face velocity: 0.4 m/s at 400 mm sash opening
- Nominal supply: 240 V / 1 ph / 50 Hz
Technical specifications - Mobility: Fully mobile unit with locking castors for secure positioning
- Docking station: Robust lockable enclosure housing water, gas, drainage and power connections with quick-release fittings for safe, repeatable connection/disconnection
- Visibility: Glass panels on all four sides for 360° viewing
- Filtration: Integral fan-driven recirculating system with a 25 kg carbon filter plus protective pre-filter; suitable for standard GCSE and A Level chemicals
- Sash: Vertically sliding toughened glass sash for user protection
- Lighting: Integral high-level lighting for demonstration visibility
- Controls: Airflow control panel for monitoring and adjustment of recirculation performance
- Electrical: RCD-protected double socket outlet; nominal supply 240 V, single phase, 50 Hz
- Services: Remote-operated gas and water outlets; drip cup, bottle trap and flexible drain hose supplied
- Safety fittings: Stainless steel restraining tether for secure installation
- Performance: Nominal face velocity 0.4 m/s at 400 mm sash opening
- Standards & testing: Designed in accordance with CLEAPSS Guide G9 and independently tested and passed by CLEAPSS
- Maintenance: Annual maintenance service available including carbon filter saturation testing and replacement in line with COSHH requirements