Centrifuge ultrafilter Vivaspin® 15 Turbo
membranefor liquidsfor protein solutions

Centrifuge ultrafilter - Vivaspin® 15 Turbo - Vivaproducts, Inc. - membrane / for liquids / for protein solutions
Centrifuge ultrafilter - Vivaspin® 15 Turbo - Vivaproducts, Inc. - membrane / for liquids / for protein solutions
Add to favorites
Compare this product

Characteristics

Filter type
membrane
Fluid
for liquids, for protein solutions, for serum
Applications
laboratory, centrifuge, for protein concentration, for protein purification
Material
polyethersulfone, PES, polypropylene
Connection type
conical
Connecting diameter

30 mm
(1.18 in)

Compressible volume

11 ml, 15 ml
(0.37 US fl oz, 0.51 US fl oz)

Description

Overview
Up to 15 ml starting volume. Concentrate up to 150x. Vivaspin® 15 Turbo centrifugal devices enable rapid concentration of protein and nanoparticle samples up to 15 ml. Designed for use in rotors accepting 50 ml centrifuge tubes, the patented twin vertical PES membrane reduces filter fouling and delivers the fastest spin times. Built-in dead stop pocket ensures maximum recoverable concentrate in a pipette-friendly format.

Applications
  • Concentration of urine prior to electrophoresis for diagnosis of multiple myeloma and amyloidosis.
  • Concentration of cerebrospinal fluid prior to electrophoresis for diagnosis of meningitis and multiple sclerosis.
  • Protein removal from serum and from cell/tissue lysates.
  • Buffer exchange for proteomic peptide samples.
  • Sample preparation for mass spectrometry and downstream analyses.


Performance — Time to concentrate up to 30x [min.] at 20°C and solute recovery %
Rotor: Swing bucket (start volume 15 ml) vs Fixed angle (start volume 11 ml). Centrifugal force: 4,000 x g (both).
Cytochrome c 0.25 mg/ml (12,400 MW) — 5,000 MWCO PES: Swing bucket Min. 30 min, Recovery 98%; Fixed angle Min. 50 min, Recovery 98%.
Lysozyme 0.25 mg/ml (14,300 MW) — 5,000 MWCO PES: Swing bucket Min. 33 min, Recovery 96%; Fixed angle Min. 50 min, Recovery 96%.
α-Chymotrypsin 1 mg/ml (25,000 MW) — 10,000 MWCO PES: Swing/Fixed Min. 10 min, Recovery 95%.
BSA 1.0 mg/ml (66,000 MW) — 10,000 MWCO PES: Swing/Fixed Min. 10 min, Recovery 99%; 30,000 MWCO PES: Swing Min. 8 min, Recovery 98%; Fixed Min. 10 min, Recovery 98%.
IgG 0.25 mg/ml (160,000 MW) — 30,000 MWCO PES: Swing Min. 23 min, Recovery 95%; Fixed Min. 17 min, Recovery 95%.

Technical Specifications
  • Concentrator capacity: Swing Bucket Rotor 15 ml; Fixed Angle Rotor 11 ml.
  • Dimensions: Total length 118 mm; Width 27 mm; Active membrane area 7.2 cm².
  • Hold-up volume (membrane and support): <10 µl.
  • Dead stop volume: Swinging bucket 100 µl; Fixed angle 60 µl.
  • Materials of construction: Body — Styrene butadiene copolymer; Filtrate vessel — Polypropylene; Concentrator cap — Polypropylene; Membrane — Polyethersulfone (PES).
  • Centrifuge compatibility: fits rotors for 50 ml (30 mm) conical bottom tubes; fixed angle minimum rotor angle 25°; maximum speed 4,000 g (devices with membrane MWCO ≥ 100 kDa require lower g forces — see data sheet).