Product description A versatile temperature-control cabinet, the KB PRO handles a wide temperature range from -20 °C to +100 °C and offers comprehensive programming options for applications beyond standard incubation. Optional light modules enable ICH photostability testing or plant-growth applications.
Benefits - Precise temperature regulation using advanced control technology
- Reliable: long-term failsafe operation ensured by proven compressor technology and a stainless-steel inner chamber
- Operational ergonomics: signal triangle with illuminated status, ergonomic control terminal with touchscreen program controller, racks with telescopic rails
- Energy-efficient operation via modern inverter cooling technology
Important characteristics - Temperature range: -20 °C to +100 °C
- APT.line™ preheating chamber technology
- Energy-efficient cooling with inverter refrigeration unit
- Climate-neutral R-290 refrigerant
- Door heating to prevent condensation
- Adjustable fan speed
- Robust housing with HIT sandwich technology
- Stainless-steel inner chamber
- Inner doors of metal-oxide-coated safety glass
- Two stainless-steel racks with telescopic rails
- 80 mm x 45 mm access port with silicone plug
- Floor-standing units with sturdy twin casters and front brakes
- Intuitive touchscreen controller with time-segment and real-time programming
- Internal data logger; measured values readable in open format via USB
- Class 3.3 independent temperature safety device (DIN 12880) with visual and acoustic alarms
- Unit self-test for comprehensive status analysis
- Data interfaces: Ethernet, USB
- BINDER Multi Management Software APT-COM™ basic edition
- Optional light modules for ICH photostability tests
- Optional LED light modules for plant growth
- Optional CO₂ gassing modules for plant growth (0–1%) and cell culture (0–20%)
Technical specifications (key data)Article Number: 9020-0434
Designation: KBPRO260-230V
Option model: Standard
Performance Data - TemperatureTemperature range: -20...100 °C
Temperature uniformity at 37°C: 0.2 ± K
Temperature fluctuation at 37°C: 0.1 ± K
Max. heat compensation at 37°C: 800 W
Recovery time after door was opened for 30 s at 37°C: 3 min
Electrical dataRated Voltage: 208...240 V
Power frequency: 50/60 Hz
Nominal power: 1,5 kW
Unit fuse: 16,0 A
Phase (Nominal voltage): 1~
Dimensions and weightsInterior volume: 262 L
Net weight of the unit (empty): 190 kg
Load per rack: 40 kg
Permitted load: 200 kg
Wall clearance back: 100 mm
Wall clearance sidewise: 180 mm
Width net: 930 mm
Height net: 1,396 mm
Depth net: 900 mm
Internal DimensionsInterior width: 656 mm
Interior height: 695 mm
Interior depth: 575 mm
Inner doors: 1
Unit doors: 1
Environment-specific dataSound-pressure level at 25°C: 48 dB(A)
Energy consumption at 25°C: 180 Wh/h
Energy consumption at 37°C: 185 Wh/h