OverviewThe laboratory fume hood is designed to protect operators and the laboratory environment from infectious aerosols and toxic gases generated during experiments. Models: FH1000(E) / FH1200(E) / FH1500(E) / FH1800(E). Suitable for chemical, biological and research laboratories.
Key features- Adjustable air speed: 9 levels; air velocity range 0.3–0.8 m/s.
- Microprocessor control with LED display and memory function for power failure.
- Manual front window: 5 mm toughened glass, height adjustable (max opening 750 mm).
- Built-in PP centrifugal blower; low noise (≤68 dB) and straightforward installation.
- Construction resistant to moderate acids and alkalis.
- Work surface height: 900 mm; worktop material: chemical-resistant phenolic resin.
- LED lighting: 20W (smaller models) or 40W (larger models).
Technical specificationsModel / External (W×D×H) / Internal (W×D×H) / System exhaust / Power consumption / Gross weight:
FH1000(E): 1000×800×2515 mm / 790×600×870 mm / 570 m³/h / 400 W / 270 kg;
FH1200(E): 1200×800×2515 mm / 990×600×870 mm / 710 m³/h / 400 W / 305 kg;
FH1500(E): 1500×800×2515 mm / 1290×600×870 mm / 930 m³/h / 500 W / 335 kg;
FH1800(E): 1800×800×2515 mm / 1590×600×870 mm / 1140 m³/h / 500 W / 365 kg;
Exhaust duct: PVC, standard length 5 m, Φ300 mm; pipe straps: 2 pcs; Noise ≤68 dB; Power supply: 220 V ±10% (50/60 Hz) or 110 V ±10% (60 Hz).
Accessories- Standard: PP water sink; base cabinet; water tap; 2 waterproof sockets (total load 500 W); 5 m exhaust duct (φ300 mm).
- Optional: air supply flow control valve; gas faucet.
Ordering and supplyBrand: BIOBASE. Origin: CHINA. Delivery time: within 7 days. Supply: factory direct. For voltage requirements, accessories or installation details contact the supplier.