IntroductionBNP32/BNP48 system is a fully-automated clinical nucleic acid extraction system that brings you walk-away automation, designed to minimize user intervention and extraction-to-extraction variability. In addition to providing quality purified eluates, the BNP32/BNP48 system offers onboard primary-sample handling and mixed-batch sampling for up to 32 or 48 samples per run.
Product detailsBrand: BIOBASE
Product origin: CHINA
Delivery time: Within 7 days
Supply capacity: Factory Direct Supply
Model: BNP32 BNP48
Features- 7-inch touch screen, easy to use, fast response
- User-defined cracking and elution temperature
- UV disinfection function, time range 1min~24hour
- Automatic control system, no need connect to computer
- Free programming to meet the needs of different reagent
- Open system, fully automatic, stable results and good repeatability
- Extract rapidly 9~40 minutes, 32/48 samples can be extracted at the same time
Technical ParametersModel: BNP32 | BNP48
Sample Quantity: 32 | 48
Processing Volume: 60µl~1000µl
Sample Volume: 20~500µl
Sample Throughput: 1~32 | 1~48
Magnetic Bead Recovery: >98%
Extracting the Difference Between Holes: CV≤3%
Heating Temperature: 8 independent heating modules, customize lysis and elution temperature(temperature range) according to your needs
Oscillating Mixing: Low, medium and high third gears are adjustable, and the fluctuation range can be adjusted with the reagent volume
Reagent Type: Magnetic bead open platform
Extraction Time: 8~60min/round (depending on the reagent used)
Internal Program: 48 groups | 50000 groups
Program Management: Powerful program editing capabilities to meet different reagent needs; U disk program import and export can be achieved
Safety Door Design: After the safety door is opened, the program operation will be automatically suspended, and the program can continue to run after the safety door closed
Built-in Air Duct: Yes
Ultraviolet Irradiation: Yes
Packing Size (W*D*H): 580*510*700 mm | 700*520*750 mm
Gross Weight: 51kg | 80kg
Caractéristiques / spécifications techniques- System type: Fully-automated clinical nucleic acid extraction system
- Capacity options: 32 samples (BNP32) or 48 samples (BNP48)
- Processing volume range: 60 µl to 1000 µl
- Sample input volume: 20 µl to 500 µl
- Magnetic bead platform: Open platform compatible with magnetic bead reagents
- Recovery performance: Magnetic bead recovery greater than 98%
- Reproducibility: CV ≤ 3% for well-to-well extraction difference
- Heating: 8 independent heating modules with customizable lysis and elution temperatures
- Mixing: Adjustable oscillating mixing (three speed levels) adaptable to reagent volume
- Automation: Free programming, no PC connection required, powerful program editing and U-disk import/export
- Safety: Program auto-suspend when safety door opened; resume after closed
- Decontamination: Built-in air duct and ultraviolet irradiation for disinfection
- Extraction time: Typical 9–40 minutes per run (8–60 min depending on reagent)
- User interface: 7-inch touchscreen
- Physical dimensions & weight: BNP32 packing 580×510×700 mm, gross weight 51 kg; BNP48 packing 700×520×750 mm, gross weight 80 kg