Product descriptionLAUDA Ultracool UC 2 W circulation chiller provides precise temperature control across an extended operating range from -10 to 35 °C with a temperature stability of ±0.5 K. The unit includes fan control for operation at low ambient temperatures, a reduced footprint compared with previous models and a standard Ethernet interface for monitoring and control. The W version features condenser cooling by water. Multiple factory options allow adaptation to customer-specific process requirements.
Key features- Industrial circulation chiller for process temperature control
- Electronic 2-point temperature control with ±0.5 K stability
- Integrated Ethernet interface (web-server operation possible)
- Compliant with EcoDesign directive 2009/125/EC energy requirements
- Graphic LCD with clear text menu navigation
- Industrial block pump: high pressure and high flow
- Stainless-steel inlet and outlet water connections
- Condenser cooling water (water-cooled version: W)
- Integrated water filter
- Suitable for non-flammable fluids (water, water/glycol)
Working rangeMin. working temperature: -10 °C
Max. working temperature: 35 °C
Temperature stability: ±0.5 K
Cooling capacity (50 Hz)25 °C: 3.1 kW
20 °C: 3.1 kW
15 °C: 2.8 kW
10 °C: 2.4 kW
5 °C: 2.0 kW
0 °C: 1.7 kW
-5 °C: 1.4 kW
-10 °C: 1.2 kW
Options / Adaptations- Speed-controlled pumps
- Flow meters
- Protection class IP 54 for outdoor installation
- Water-cooled version (W)
Technical specifications- Working temperature range: -10 … 35 °C
- Ambient temperature range: -15 … 50 °C
- Temperature stability: ±0.5 K
- Max. filling volume: 12 L
- Max. discharge pressure: 3.4 bar
- Max. pump flow rate: 42 L/min
- Nominal pump flow rate: 5.6 L/min
- Inlet / Outlet connection thread (inside): Rp 1/2
- Water cooling connection thread (inside): Rp 1/2
- Dimensions (W×D×H): 510 x 680 x 1042 mm
- Weight: 89 kg
- Noise level: 53.5 dB(A)
- Power supply: 230 V, 50 Hz
- Power plug: Power cord without plug (HAR)