DescriptionThe Kaye ValProbe RT temperature data loggers with bendable sensors measure temperatures from -85°C to 400°C. Available in single-, dual- and five-channel bendable sensor configurations, they combine precision platinum RTD sensing with a low-power electronic design for extended battery life. Operator-programmable sample rates begin at 1 second and each sensor stores up to 100,000 data points. Antenna diversity provides robust RF communication for real-time data retrieval.
Key features- Precision platinum RTD sensing element (PT-1000)
- Operator-programmable sampling from 1 s; up to 100,000 samples per sensor
- Field-replaceable 3.6V lithium battery
- Antenna diversity for reliable RF communications
- Logger housing: stainless steel 316L and PEEK
1- and 2-Channel Bendable Logger- Sensor measurement range: -85°C to 400°C
- Available as single- and dual-sensor configurations
- Sensor lengths: 12", 24", 36" (305, 610, 915 mm)
- Sensor tip diameter: 0.095" (2.4 mm)
5-Channel Bendable Logger- Sensor measurement range: 0°C to 400°C
- Five bendable sensors
- Sensor length: 24" (610 mm)
- Sensor tip diameter: 0.078" (2.0 mm)
Applications- Steam sterilizers (autoclaves)
- Dry‑heat sterilizers
- Steam‑in‑place (SIP)
- Water cascade / fall sterilizers
- Incubators
- Stability chambers
- Freezers
- Freeze drying / lyophilization
- Process vessels and tanks
Specifications (summary)Sensor type — Single & dual sensors (1/2‑channel) — Five sensors (5‑channel)
Sensor length — 12", 24", 36" (305, 610, 915 mm) — 24" (610 mm)
Tip diameter — 0.095" (2.4 mm) — 0.078" (2.0 mm)
Measurement range — -85°C to 400°C — 0°C to 400°C
Accuracy — -85°C to 140°C: ±0.1°C; 140°C to 400°C: ±0.25°C — 0°C to 140°C: ±0.1°C; 140°C to 400°C: ±0.25°C
Technical specifications- Logger dimensions: Height 1.9" / Diameter 1.4" (48 mm / 36 mm)
- Logger material: Stainless Steel 316L and PEEK
- Battery: Field-replaceable 3.6V lithium
- Sampling rate: Operator-programmable, starting at 1 second
- Data storage: Up to 100,000 samples per sensor in non‑volatile memory
- Real‑time clock accuracy: < 15 seconds/day
- Calibration: NVLAP (NVLAP LAB CODE 200913-0) / DAkkS calibration available
- Verification: Automated user verification capability
- Sensing element: Precision platinum RTD (PT-1000)
- Environmental (logger): -85°C to 140°C
- Environmental pressure: 0–5 bar absolute
- Environmental humidity: 0–100% condensing
- Regulatory compliance: UL and CE