Pressure data-logger ValProbe
temperaturesterilizationlaboratory

Pressure data-logger - ValProbe - Kaye - temperature / sterilization / laboratory
Pressure data-logger - ValProbe - Kaye - temperature / sterilization / laboratory
Pressure data-logger - ValProbe - Kaye - temperature / sterilization / laboratory - image - 2
Pressure data-logger - ValProbe - Kaye - temperature / sterilization / laboratory - image - 3
Add to favorites
Compare this product

Characteristics

Data type
pressure, temperature
Applications
sterilization, laboratory, for temperature traceability
Other characteristics
battery-powered
Temperature

Max.: 140 °C
(284 °F)

Min.: 0 °C
(32 °F)

Diameter

46 mm
(1.8 in)

Description

Description
The combined pressure/temperature Kaye ValProbe data logger provides an integrated solution for small-scale sterilizer validation, monitoring and routine control. It is suited for benchtop and laboratory sterilizers, hospital validation, dental offices and other medical-device sterilization workflows.

Features
  • Combined pressure and temperature sensing in a single compact logger
  • High-accuracy measurements to support regulatory validation
  • 10,000-sample memory (5000 samples per sensor stored in EEPROM)
  • Field-replaceable 3.6V lithium battery
  • Programmable sample interval from 1 second to 12 hours
  • Configurable start, change and stop event triggers
  • Designed to enable compliance with ISO 17665, EN 554, EN 285 and HTM 2010


Applications
  • Steam sterilizer validation and process qualification
  • Hospital sterilization validation and routine process control
  • Small-scale sterilization such as benchtop sterilizers in dental and clinical settings
  • Parametric release and sterilization record-keeping


Specifications (Sensing Element - Temperature)
Sensing element: Pressure Platinum RTD
Measurement range and accuracy: 0°C to 140°C ±0.1°C
Environmental range: -45°C to 140°C; 0–100% RH, condensing
Pressure environment: 0 to 10 bar absolute (0 to 147 psia)
Logger material: 316 stainless steel
Battery: Field-replaceable 3.6V lithium
Sampling rate: 1 second to 12 hours
Data storage: 5000 samples per sensor retained in EEPROM (product specifications indicate 10,000 sample memory)
Calibration: Factory calibrated (NIST-traceable) with user calibration option
Real-time clock accuracy: 15 seconds per 24 hours (0.0174%) from -45°C to 140°C
Regulatory compliance: UL, CE, Intrinsically Safe

Specifications (Sensing Element - Pressure)
Sensing element: Absolute pressure sensor
Operating range: 0 to 5 bar absolute (0 to 73 psia)
Accuracy: >0°C to 120°C ±25 mbar; 120°C to 135°C ±10 mbar; 135°C to 140°C ±25 mbar
Logger dimensions: Diameter 46 mm x 35 mm; Height 72 mm
Calibration: Factory calibrated (NIST-traceable / DAkkS-traceable) with user calibration option
Part number: x2532

Technical specifications
  • Product: Kaye ValProbe Pressure/Temperature data logger
  • Temperature range: 0°C to 140°C ±0.1°C
  • Environmental range: -45°C to 140°C; 0–100% RH condensing
  • Pressure range: 0 to 5 bar absolute (0 to 73 psia)
  • Pressure accuracy: ±25 mbar (0–120°C), ±10 mbar (120–135°C), ±25 mbar (135–140°C)
  • Material: 316 stainless steel
  • Battery: Field-replaceable 3.6V lithium
  • Data storage: 5000 samples per sensor retained in EEPROM
  • Sampling rate: Programmable 1 s to 12 h
  • Real-time clock accuracy: 15 s per 24 h across -45 to 140°C
  • Regulatory: UL, CE, Intrinsically Safe
  • Dimensions: Ø46×35 mm; Height 72 mm
  • Part number: x2532

Other Kaye products

Validation Systems

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.
About MedicalExpo
MedicalExpo is a marketplace that offers B2B players a complete experience, from product sourcing to purchase. MedicalExpo is an entity of VirtualExpo Group.