OverviewThe Kaye ValProbe RT humidity/temperature data logger uses RF technology to deliver real-time measurements to the Validation Console. Measurement ranges: humidity 15%–95% RH; temperature 0°C–70°C. Digital sensor technology and optimized electronics provide accurate readings and extended battery life. Sampling is operator-programmable from 1 second and each sensor can record up to 100,000 data points in non-volatile memory.
Field-replaceable sensorThe ValProbe RT features a field-replaceable digital RH/T sensor to enable fast, cost-effective replacement without returning the entire logger. Sensors are shipped pre-calibrated to ensure measurement traceability and consistency.
Easy verifications in your labTemperature verification is simplified by attaching the 1 m sensor extension cable and using a dry-block calibrator (e.g. Kaye LTR-150) to validate multiple loggers simultaneously. The ValProbe RT software supports automated multi-point temperature checks. For relative humidity verification, portable humidity test chambers allow simultaneous evaluation of several ValProbes; the software supports automated humidity verification and automatic report generation.
Features- Accuracy: ±0.15°C (0°C–70°C) for temperature; ±2% RH (15%–95% RH) for humidity
- 100,000 samples per sensor stored in non-volatile memory
- Field-replaceable digital temperature/humidity sensor
- 1 m sensor extension cable
- Field-replaceable 3.6V Lithium battery
- Operator-programmable sample rate, start, delay and stop functions (from 1 s)
Applications- Stability chambers
- Incubators
- Warehouses and storage areas
- Temperature chambers
Technical specifications- Product type: Humidity/Temperature data logger
- Model / Series: ValProbe RT
- Sensor type: Digital RH/T sensor (field-replaceable)
- Sensor length: Up to 1 m (extension cable available)
- Sensor tip diameter: 0.295" (7 mm); outer filter 0.472" (12 mm)
- Measurement range: Temperature 0°C to 70°C; Humidity 15% to 95% RH
- Accuracy: Temperature ±0.15°C; Humidity ±2% RH
- Logger dimensions: Height ~48 mm / Diameter ~36 mm
- Material: Stainless steel 316L and PEEK
- Battery: Field replaceable 3.6V Lithium
- Sampling rate: From 1 second (programmable)
- Data storage: 100,000 samples per sensor in non-volatile memory
- Real-time clock accuracy: < 15 sec/day
- Calibration: NVLAP (LAB CODE 200913-0) / DAkkS available
- Verification: Automated user verification via software
- Operating conditions: 0°C to 70°C; non-condensing humidity
- Regulatory: UL and CE