OverviewAnalytical balance with Single-Cell weighing system, model ACJ 80-4M. Optional conformity assessment (calibration) ex works.
Key features- Automatic internal adjustment when temperature change ≥ 2 °C or time-controlled every 4 h
- Dosage aid: high stability mode and selectable filter settings
- Recipe weighing and documentation with combined tare/print function; ingredients numbered and printed with number and nominal weight
- Automatic data output to PC/printer each time the balance is steady
- Programmable 4-digit identification number printed on calibration protocol
- Protective working cover included
Applications- Pharmacies
- Food analysis
- Environmental analysis
- Chemical industry
- Research and development
Technical data (summary)Weighing capacity [Max]: 82 g
Readability [d]: 0.0001 g
Verification scale interval [e]: 0.001 g
Units: g, mg
Minimum weight [Min]: 0.01 g
Linearity: ± 0.0003 g
Reproducibility: 0.0002 g
Stabilization time: 3 s
Technical specifications- Model: ACJ 80-4M
- Weighing capacity (Max): 82 g
- Readability (d): 0.0001 g
- Verification scale interval (e): 0.001 g
- Units: g, mg
- Minimum weight (Min): 0.01 g
- Linearity: ± 0.0003 g
- Reproducibility: 0.0002 g
- Stabilization time: 3 s
- Tolerance weighing: yes (visual)
- Interfaces: RS-232 (standard), USB-Device
- Adjustment options: internal calibration (automatic), internal weight (manual)
- Recommended adjustment weight: 80 g (E2)
- Ambient temperature: 10 °C – 30 °C
- Weighing pan diameter (Ø): 91 mm
- Draft shield material: glass
- Housing material: cast aluminium, plastic
- Housing dimensions (W×D×H): 213×333×338 mm
- Weighing plate material: stainless steel
- Net weight approx.: 7 kg
- Construction type: single-range balance
- GTIN / EAN: 4045761235670