Tomography test phantom VD1003110
for CT scanheadtorso

Tomography test phantom - VD1003110  - Orion France - for CT scan / head / torso
Tomography test phantom - VD1003110  - Orion France - for CT scan / head / torso
Add to favorites
Compare this product

Characteristics

Type of calibration
for tomography, for CT scan
Area of the body
torso, head

Description

Summary
The CTDI Head and Body Phantom in PMMA is a cylindrical set designed for measuring the computed tomography dose index (CTDI) and dose-length product (DLP) on CT systems. Compliant with IEC 61223-3-5 and IEC 61223-2-6, it enables dosimetric performance assessment and constancy checks of clinical CT protocols.

Composition and modules
  • Two cylindrical modules: head and body.
  • Head module: external diameter 160 mm (16 cm), thickness 150 mm.
  • Body module: external diameter 320 mm (32 cm), thickness 150 mm.

Holes and accessories
Each module includes apertures for CTDI-100 ionization chambers: one central hole and four peripheral holes arranged at 90°, located 10 mm from the outer edge. Acrylic plugs are supplied to close unused cavities. Engraved alignment axes facilitate centering on the CT table.

Applications
This phantom is used for CT dose measurements (CTDI and DLP), protocol validation, constancy checks of equipment and reproducibility studies of patient doses.

Models
  • VD1003110 - CT Dose Measurement Head and Body Phantom
  • VD0225250 - Transport case with fitted interior

Characteristics / technical specifications
  • Compliance: IEC 61223-3-5 and IEC 61223-2-6.
  • Head phantom: external diameter 160 mm, thickness 150 mm, 5 cavities (1 central + 4 peripheral).
  • Body phantom: external diameter 320 mm, thickness 150 mm, 5 cavities (1 central + 4 peripheral).
  • Holes for CTDI-100 chambers: one central and four peripheral at 90°, positioned 10 mm from the outer edge.
  • Acrylic plugs supplied to close unused cavities.
  • Centering aids: engraved alignment axes on modules.
  • Included: 9 acrylic inserts (listed in the technical specifications).
  • Applications: CTDI/DLP measurement, protocol validation, constancy control and dose reproducibility studies.

Catalogs

No catalogs are available for this product.

See all of Orion France‘s catalogs

Other Orion France products

Products

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.