Tomography test phantom VD1003105
for CT scanheadtorso

Tomography test phantom - VD1003105  - Orion France - for CT scan / head / torso
Tomography test phantom - VD1003105  - Orion France - for CT scan / head / torso
Add to favorites
Compare this product

Characteristics

Type of calibration
for tomography, for CT scan
Area of the body
torso, head
Patient type
pediatric

Description

Description The CT phantom PMMA 3-part VD1003105 is designed for CTDI measurement and routine quality control of CT scanners. Manufactured from high-density transparent PMMA, the phantom consists of three stackable cylinders representing a pediatric head (Ø 10 cm), an adult head / pediatric body (Ø 16 cm) and an adult body (Ø 32 cm). Each section includes measurement holes for insertion of a pencil-type ionization chamber (e.g. IBA DCT10-RS) to obtain accurate and repeatable dose readings. The phantom complies with FDA 21 CFR 1020.33 and is suitable for acceptance testing, protocol verification and performance monitoring of CT systems.

Models
  • VD1003105 - CT phantom PMMA 3-part — head, body, pediatric
  • VD0225250 - Transport case


Features
  • Material: high-density transparent PMMA
  • Configuration: modular 3-piece stackable assembly (head/body/pediatric)
  • Cylinder diameters: Ø 10 cm – Ø 16 cm – Ø 32 cm
  • Measurement cavities: 5 holes (10 cm) – 4 holes (16 cm and 32 cm)
  • Filling plugs: 13 acrylic plugs included
  • Instrumentation compatibility: pencil-type ionization chambers (e.g. IBA DCT10-RS)
  • Regulatory compliance: FDA 21 CFR 1020.33 (CT systems)
  • Applications: CTDI measurement, quality control, constancy tests and equipment acceptance


Technical specifications
  • Material: high-density transparent PMMA
  • Modular configuration: 3 stackable parts (pediatric head, adult head / pediatric body, adult body)
  • Diameters: 10 cm (pediatric head), 16 cm (adult head / pediatric body), 32 cm (adult body)
  • Measurement ports: pre-drilled holes for insertion of pencil-type ionization chamber
  • Number of measurement cavities: 5 (10 cm), 4 (16 cm and 32 cm)
  • Included accessories: 13 acrylic filling plugs
  • Compatible instruments: pencil-type ionization chambers (e.g. IBA DCT10-RS)
  • Regulatory: FDA 21 CFR 1020.33 (CT systems)
  • Intended use: CTDI measurement, QA, acceptance testing, clinical protocol verification and performance follow-up