Product descriptionThe regional gas pressure alarm monitors pressure in gas supply pipelines for areas such as hospital wards, nurse stations, operating rooms, ICU gas supply manifolds and gas valve boxes. It provides real-time pressure indication via a 0.8-inch digital tube. The pressure sensor connects to the pipeline and outputs a digital carrier signal; the system converts this signal to indicate pressure changes and trigger alarms.
Technical tableSurface mounted, high-precision 1.6 level electric contact pressure gauge
Number of pressure monitoring points: 1–6 channels
Detection range: 0.1 MPa–1 MPa (adjustable)
Alarm method: sound and light warning
Buzzer mute: default silent recovery time 10 minutes; alarm restores after detecting a new abnormality
Host power supply voltage: DC24V ± 6%
Can be customized for multiple gases
Characteristics / Technical specifications- Surface mounted, high-precision 1.6 level electric contact pressure gauge
- Number of pressure monitoring points: 1–6 channels
- Detection range: 0.1 MPa–1 MPa (adjustable)
- Alarm method: sound and light warning
- Buzzer mute: default silent recovery time 10 minutes; alarm restores after detecting a new abnormality
- Host power supply voltage: DC24V ± 6%
- Customizable for multiple gases
- Real-time pressure display via 0.8-inch digital tube
- Pressure sensor connects to gas pipeline and outputs a digital carrier signal