Product OverviewLabotemp Asphalt Ignition Furnace LBC-101 is a precision testing system designed for accurate determination of asphalt binder content in hot mix asphalt samples using the ignition method. The system eliminates the need for solvent extraction by providing a faster, safer, and more environmentally responsible alternative. It features an integrated high-accuracy balance that continuously monitors sample weight during combustion and automatically calculates asphalt binder content with up to 0.10% accuracy. The furnace supports a maximum sample capacity of 4000 g, includes intelligent endpoint detection, automated operation, and real-time data monitoring via a large industrial touch-screen interface.
Key Capabilities- Precision asphalt binder analysis with 0.10% test accuracy
- Maximum sample capacity: 4000 g; recommended sample weight: 1000–2000 g
- Integrated precision balance for continuous weight monitoring and automatic calculations
- Standard operating temperature 538°C with maximum chamber temperature up to 800°C and secondary combustion chamber up to 900°C
- Automated, one-touch operation with automatic endpoint detection and automatic printing of results
Features- Real-time monitoring (temperature, sample mass, weight loss, correction factor, asphalt content) via industrial 7-inch touch-screen interface
- Complies with AASHTO T308, ASTM D6307, and NCAT ignition method requirements
- Intelligent temperature control system for stable, repeatable combustion conditions; temperature accuracy ±5°C
- Balance capacity 12 kg with balance accuracy 0.1 g
- Secondary combustion chamber with exhaust filtration and secondary combustion filter device to reduce emissions
- Built-in self-diagnostic and alarm functions for operational reliability and user safety
- High-performance thermal insulation keeping external surface temperature < 45°C
- Flexible data processing and correction factor settings to support different aggregate and asphalt mixture characteristics
- Designed for continuous operation in asphalt plants, highway laboratories, research facilities, and quality control environments
Applications- Determination of asphalt binder content in hot mix asphalt (HMA) for mix design verification and production control
- Compliance testing in highway laboratories, transportation agencies, and accredited materials testing facilities
- Batch validation, process optimization, and consistency monitoring at asphalt plants
- Quality control, research studies on aggregate–binder ratios, material performance and pavement durability
- Routine acceptance testing, specification compliance, and forensic pavement investigations in civil engineering laboratories
Standard Accessories- Sample baskets
- Basket covers
- Trays
- Basket retainer brackets
- Rolls of printer paper
- Transport handle
- Aluminum cooldown plate
- Cool down safety cage
- Fan motor lubricant
- Face shield
Caractéristiques / Spécifications techniques- Maximum Temperature: 800°C
- Standard Working Temperature: 538°C
- Balance Capacity: 12 kg
- Power Supply: AC 380V 50Hz ±10V, 15A
- Maximum Sample Weight: 4000 g
- Recommended Sample Weight: 1000–2000 g
- Temperature Accuracy: ±5°C
- Secondary Combustion Chamber Temperature: 900°C
- Balance Accuracy: 0.1 g
- Asphalt Content Test Accuracy: 0.10%
- Combustion Chamber Dimensions (L × W × H): 350 × 440 × 330 mm
- Operation Mode: Automatic
- Result Printing: Automatic printing of test results
- Thermal Insulation: External surface temperature < 45°C
- Exhaust Treatment: Secondary combustion filter device
- Gross Weight: 260 kg
- Net Weight: 210 kg
- Overall Dimensions (L × W × H): 830 × 660 × 1780 mm
- Package Size: 1000 × 770 × 1680 mm