IntroductionThe HSIT-01 half-suit integrity tester is designed to verify the integrity of a half-suit installed on isolators used in aseptic processes. It performs controlled filling and pressure monitoring to detect leaks or assembly defects before production start.
Features- High-sensitivity pressure sensors monitor filling pressure and trigger automatic stop at preset values.
- Emergency stop protects the half-suit and equipment from overpressure or abnormal conditions.
- Rotating port facilitates alignment and connection to the half-suit port.
- Mobile low-profile design allows the unit to slide under the half-suit for in-situ testing.
- PLC control with touchscreen ensures clear operator interface and configurable test parameters.
Applications- Pharmaceutical and biotechnology aseptic production isolator testing
- Food industry hygiene and containment checks
- High-risk industrial protection systems verification
- Cleanroom and isolator commissioning and maintenance
Technical specifications- Model: HSIT-01
- Function: Integrity check of half-suits installed on isolators
- Pressure monitoring: High-sensitivity sensors with automatic stop at preset pressure
- Safety: Emergency stop and automatic shutdown on preset limits
- Mechanics: Rotating half-suit port for accurate alignment
- Design: Mobile unit with low profile to slide under the half-suit
- Control: PLC control with touchscreen for parameter setting and test logging
- Operation: Automatic stop on reaching preset pressure to prevent damage to the half-suit