The kit has been designed for the quantitative determination of β2 Microglobulin (β2-MG) in human serum.
The method can be used for samples over the range of 0.1~10.0 μg/mL.
Product Description
PRINCIPLE OF THE TEST:
Sandwich method:
β2-MG Antibody labeled by FITC and β2-MG Antibody pair labeled by AP bind with the β2-MG antigen in the sample, control or calibrator and form sandwich complex. Then add the magnetic beads binding with Anti-FITC, through the specific binding between the FITC and Anti-FITC, the complex will be bound by the magnetic beads. Then the entire complex will be captured by the external applied magnetic field and separate from the unbound substance. After 3-step washing, add the substrate. The substrate will be catalytically cracked under the action of the enzyme, and form an unstable intermediate in excited state. When the intermediate in excited state returns to the ground state, it will issue photons and make a light-emitting reaction. Then the CLIA analyzer will measure the luminous intensity and count the results through software by comparing the luminous intensity with the cutoff value to determine whether the corresponding antibody exists.
DESCRIPTION:
Product name - β2-MG Diagnostic Reagent
Luminous principle - Enzymatic chemiluminescence
Luminous Marker - AP(alkaline phosphatase)
Specification - 100 test/kit
50 test/kit
Principle - Sandwich Method
Components - Magnetic beads
Anti-A/Anti-B
Calibrators
QC 1
QC 2
Sample - Serum
Storage - 2-8℃