The kit has been designed for the quantitative determination of Alpha-Fetoprotein (AFP) in human serum.
The method can be used for samples over the range of 0.5-1200.0 IU/mL.
The test has to be performed Taizhou Zecen Biotech Co., Ltd. model CIA 600, CIA 600Plus, CIA
1200, CIA 1200M, CIA 1800, CIA 2800 and POClia 8, POClia minus, POClia plus, POClia auto
automatic chemiluminescence analyzer. Shenzhen IncreCare Biotech Co., Ltd. model Shine i1900、
Shine i2900 automatic chemiluminescence analyzer.
Product Description
Storage and Stability
⚫ Sealed: Stored at 2-8℃ until the expiry date.
⚫ Opened: Stable for 4 weeks. To ensure best kit’s performance, it is recommended to place opened kits in the refrigerator if it’s not going to be used on board during next 12 hours.
⚫Keep upright.
⚫Keep away from sunlight.
SPECIMEN COLLECTION AND PREPARATION
Sample material: serum
Collect samples using standard procedures.
Store at 2-8℃: 24 hours, for longer storage periods: freeze to below - 20℃
Avoid repeated freezing and thawing cycles, stored samples should be thoroughly mixed prior to use (Vortex mixer). Please ask local representative of ZECEN for more details if you have any doubt.
Shipping
• Before shipping specimens, it is recommended that specimens be removed from the serum or plasma separator, red blood cells or clot. When shipped, specimens must be packaged and labeled in compliance with applicable state, federal and international regulations covering the transport of clinical specimens and infectious substances. Specimens must be shipped frozen (dry ice). Do not exceed the storage time limitations identified in this section of the package insert.