Benefits- Uniform drying conditions thanks to APT.line™ preheating chamber
- Identical test conditions throughout the usable chamber independent of sample size and quantity
- Excellent thermal insulation reduces operating costs
Product descriptionThe BINDER ED 400 drying and heating chamber is designed for routine drying and sterilization tasks up to +300 °C. Natural convection ensures efficient thermal processes and fast, uniform drying across the 400 L usable volume. The unit is suitable for industrial and laboratory applications requiring repeatable temperature conditions.
Important characteristics- Temperature range: from +5 °C above ambient temperature to +300 °C
- APT.line™ preheating chamber technology for homogeneous temperature distribution
- Natural convection
- Adjustable exhaust air flap
- Controller with timer functions
- 2 chrome-plated racks (standard); up to 10 shelf positions possible
- Class 2 independent adjustable temperature safety device (DIN 12880) with visual alarm
Technical specificationsDataArticle Number: 9010-0075 | 9010-0168
Designation: ED400-400V-I | ED400UL-208V-I
Option model: with RS 422
Performance Data TemperatureTemperature range: +5 °C above ambient temperature to 300 °C
Temperature uniformity at 150 °C: 3.0 ± K
Temperature fluctuation at 150 °C: 0.5 ± K
Heating up time to 150 °C: 71 min
Recovery time after door was opened for 30 s at 150 °C: 31 min
Air change dataAir change rate at 150 °C: 10 x/h
Electrical dataRated Voltage: 400 V (variant) | 208 V (variant)
Power frequency: 50/60 Hz (400 V variant) | 60 Hz (208 V variant)
Nominal power: 3.4 kW
Unit fuse: 16.0 A
Phase (Nominal voltage): 3~/N/PE (400 V variant) | 3~ (208 V variant)
Dimensions and weightsInterior volume: 400 L
Net weight (empty): 125 kg
Load per rack: 35 kg
Permitted load: 90 kg
Wall clearance back: 160 mm
Wall clearance sidewise: 100 mm
Housing dimensions (net, not incl. fittings and connections)Width net: 1,235 mm
Height net: 1,025 mm
Depth net: 765 mm
Internal DimensionsInterior width: 1,000 mm
Interior height: 800 mm
Interior depth: 520 mm
Unit doors: 2
Environment-specific dataEnergy consumption at 150 °C: 672 Wh/h
FixturesNumber of shelves (std./max.): 2/10
Technical characteristics / specifications- Model family: ED series (natural convection drying and heating chambers)
- Usable interior volume: 400 L
- Working temperature up to +300 °C
- APT.line™ technology for uniform preheating
- Temperature accuracy: uniformity 3.0 ± K at 150 °C, fluctuation 0.5 ± K at 150 °C
- Heating time to 150 °C: approx. 71 min; recovery after 30 s door opening at 150 °C: approx. 31 min
- Air change rate at 150 °C: 10 exchanges/hour
- Electrical: nominal power 3.4 kW; rated voltage variants 400 V / 208 V; unit fuse 16 A
- Net weight (empty): 125 kg; permitted load 90 kg; load per rack 35 kg
- Exterior net dimensions (W×H×D): 1,235 × 1,025 × 765 mm; interior (W×H×D): 1,000 × 800 × 520 mm
- Standard equipment: 2 chrome-plated racks; up to 10 shelves possible (2/10 std./max.)
- Safety: Class 2 independent adjustable temperature safety device (DIN 12880) with visual alarm