Heating oven ED-S 56
dryingfor scientific researchlaboratory

Heating oven - ED-S 56 - BINDER GmbH - drying / for scientific research / laboratory
Heating oven - ED-S 56 - BINDER GmbH - drying / for scientific research / laboratory
Heating oven - ED-S 56 - BINDER GmbH - drying / for scientific research / laboratory - image - 2
Add to favorites
Compare this product

Characteristics

Function
heating, drying
Applications
laboratory, for scientific research, for general purposes
Type
natural convection
Temperature control
with ventilation system
Configuration
benchtop, compact
Number of doors
1-door
Temperature range

Max.: 250 °C
(482 °F)

Min.: 7 °C
(44.6 °F)

Capacity

62 l
(16.38 gal)

Description

Product description
BINDER Solid.Line ED-S 56 is a compact drying and heating chamber designed for reliable drying and warm-storage tasks in laboratories, research and quality assurance. The appliance combines natural convection with APT.line™ preheating chamber technology to provide reproducible temperature conditions.

Benefits
  • Easy handling and ergonomic door operation
  • Precise temperature control with timer function
  • Adjustable exhaust flap for controlled atmosphere


Key characteristics
  • Temperature range: ambient +7 °C to +250 °C
  • APT.line™ preheating chamber technology
  • Natural convection
  • Adjustable exhaust air flap
  • Controller with timer function
  • 1 chrome-plated rack included with shelf supports
  • Class 2 integrated independent adjustable temperature safety device (DIN 12880) with visual alarm


Technical specifications (selected)
Data:
Article Number: 9090-0014 (230 V) / 9090-0015 (120 V)
Designation: EDS056-230V / EDS056UL-120V
Model: ED-S 56 (Standard)

Performance — Temperature:
Temperature range: +7 °C above ambient temperature to 250 °C
Temperature uniformity at 150 °C: 3.4 ± K
Temperature fluctuation at 150 °C: 0.8 ± K
Heating up time to 150 °C: 60 min
Recovery time after door was opened for 30 s at 150 °C: 45 min

Electrical data:
Rated voltage: 230 V (article 9090-0014) / 120 V (article 9090-0015)
Power frequency: 50/60 Hz (230 V model) / 60 Hz (120 V model)
Nominal power: 1.05 kW (230 V) / 1.15 kW (120 V)
Unit fuse: 6.3 A
Phase (nominal voltage): 1~

Dimensions and weights:
Interior volume: 62 L
Net weight (empty): 36 kg
Load per rack: 15 kg
Permitted total load: 45 kg
Wall clearance back: 160 mm
Wall clearance sidewise: 100 mm

Housing dimensions (not incl. fittings and connections):
Width net: 614 mm
Height net: 625 mm
Depth net: 577 mm

Internal dimensions:
Interior width: 400 mm
Interior height: 440 mm
Interior depth: 350 mm
Unit doors: 1

Environment-specific data:
Energy consumption at 150 °C: 210 Wh/h

Fixtures:
Number of shelves (std./max.): 1/3

Characteristics / technical specifications
  • Model: ED-S 56
  • Article numbers: 9090-0014 (230 V), 9090-0015 (120 V)
  • Designation: EDS056-230V / EDS056UL-120V
  • Temperature range: ambient +7 °C to 250 °C
  • Interior volume: 62 L
  • Interior dimensions (W×H×D): 400 × 440 × 350 mm
  • Net weight (empty): 36 kg
  • Nominal power: 1.05 kW (230 V) / 1.15 kW (120 V)
  • Rated voltage: 230 V or 120 V
  • Power frequency: 50/60 Hz (230 V) / 60 Hz (120 V)
  • Energy consumption at 150 °C: 210 Wh/h
  • Number of shelves (std./max.): 1/3 (chrome-plated wire shelf available as accessory)
  • Permitted total load: 45 kg

Exhibitions

Meet this supplier at the following exhibition(s):

Lab-Supply Dresden