Product descriptionA BINDER material test chamber of the FP series delivers consistently short drying times and rapid heating. Additional program functions and an adjustable fan allow precise control of temperature and convection conditions even under full load and during extended runs.
Benefits- High temperature accuracy thanks to APT.line™ technology
- User-friendly controller with large display
- Computer interface: Ethernet
- Independent adjustable temperature safety device Class 2 (DIN 12880) and selectable Class 3.1 safety device with visual/acoustic alarm
Important characteristics- Temperature range: +12 °C above ambient up to +300 °C
- High temperature accuracy with APT.line™ technology
- Adjustable fan speed
- Controller with interval and real-time programming
- Electromechanical exhaust air flap control
- 2 chrome-plated racks (standard)
- Ergonomic handle design
- Internal data logger; measured values exportable in open format via USB
Data (article numbers and designation)Article Number: 9010-0366 | 9010-0370
Designation: FP720-400V | FP720UL-208V
Option model: Standard | Standard
Performance Data TemperatureArticle Number: 9010-0366 | 9010-0370
Temperature range: +12 °C above ambient temperature to 300 °C
Temperature uniformity at 150 °C: 2.4 ± K
Temperature fluctuation at 150 °C: 0.6 ± K
Heating up time to 150 °C: 25 min
Recovery time after door was opened for 30 s at 150 °C: 6 min
Air change dataArticle Number: 9010-0366 | 9010-0370
Air change rate at 100 °C: 5 x/h
Electrical dataArticle Number: 9010-0366 | 9010-0370
Rated Voltage: 400 V | 208 V
Power frequency: 50/60 Hz | 60 Hz
Nominal power: 4.5 kW
Unit fuse: 16.0 A
Phase (Nominal voltage): 3~/N/PE | 3~
Dimensions and weightsArticle Number: 9010-0366 | 9010-0370
Interior volume: 741 L
Net weight of the unit (empty): 166 kg
Load per rack: 45 kg
Permitted load (total): 315 kg
Wall clearance back: 160 mm
Wall clearance sidewise: 100 mm
Housing dimensions not incl. fittings and connectionsArticle Number: 9010-0366 | 9010-0370
Width net: 1,160 mm
Height net: 1,590 mm
Depth net: 816 mm
Internal DimensionsArticle Number: 9010-0366 | 9010-0370
Interior width: 1,000 mm
Interior height: 1,300 mm
Interior depth: 574 mm
Unit doors: 2
Environment-specific dataArticle Number: 9010-0366 | 9010-0370
Sound-pressure level: 43 dB(A)
Energy consumption at 100 °C: 510 Wh/h
Energy consumption at 150 °C: 880 Wh/h
Accessories & Options (selection)- Interior lighting (two 27 W bulbs) – item 8012-2191
- Access ports with silicone plug (various diameters and positions) – e.g. 100 mm left 8012-2207, 100 mm right 8012-2209, 10/30/50 mm variants
- HEPA air filter for fresh air supply (H14) – item 8012-2222
- Inner chamber, reinforced (max total load 350 kg) – item 8012-2226
- Door lock (lockable door handle) – item 8012-2238
- Analog output 4-20 mA for temperature values – item 8012-2188
- Additional Pt100 temperature sensor (interior) – item 8012-2193
- Various racks and shelves (chrome-plated, stainless steel, perforated, heavy-load) – e.g. 8012-2044, 8012-2170, 8012-2178, 8012-2185
- Software APT-COM™ (BASIC / PROFESSIONAL / GLP editions) for logging and management – e.g. 9053-0039, 9053-0040, 9053-0042
- Calibration certificates and extended measurement services (9 / 18 / 27 measuring points etc.) – items 8012-2227 to 2232 and DL30-/DL33- prefixed services
- Service contracts and commissioning options (maintenance contracts BRONZE/SILVER/GOLD, commissioning, IQ/OQ/PQ execution) – various DLxx codes
Technical specifications- Model name: FP 720
- Temperature range: +12 °C above ambient to 300 °C
- Temperature uniformity (150 °C): 2.4 ± K
- Temperature fluctuation (150 °C): 0.6 ± K
- Heating time to 150 °C: 25 min
- Recovery time after 30 s door opening at 150 °C: 6 min
- Air change rate at 100 °C: 5 x/h
- Rated voltage / frequency: 400 V / 50/60 Hz (standard) or 208 V / 60 Hz (alternative)
- Nominal power: 4.5 kW
- Unit fuse: 16 A
- Phasing: 3~/N/PE (400 V) or 3~ (208 V)
- Interior volume: 741 L
- Interior dimensions (W×H×D): 1,000 × 1,300 × 574 mm
- Net weight (empty): 166 kg
- Load per rack: 45 kg (2 racks standard, up to 16 slots)
- Permitted total load: 315 kg
- Doors: 2
- Sound-pressure level: 43 dB(A)
- Energy consumption (100 °C / 150 °C): 510 Wh/h / 880 Wh/h
- Interfaces: Ethernet, USB (for internal data logger export)
- Safety devices: Class 2 independent adjustable temperature safety device (DIN 12880) and Class 3.1 independent temperature safety device available
- Special features: APT.line™ technology for high temperature accuracy, adjustable fan speed, controller with interval and real-time programming, electromechanical exhaust flap control